Improving the energy resolution of bent crystal X-ray spectrometers with position-sensitive detectors
Wavelength-dispersive high-resolution X-ray spectrometers often employ elastically bent crystals for the wavelength analysis. In a preceding paper [Honkanen et al. (2014). J. Synchrotron Rad. 21, 104-110] a theory for quantifying the internal stress of a macroscopically large spherically curved anal...
Veröffentlicht in: | Journal of synchrotron radiation. - 1994. - 21(2014), Pt 4 vom: 21. Juli, Seite 762-7 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , , , |
Format: | Online-Aufsatz |
Sprache: | English |
Veröffentlicht: |
2014
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article Research Support, Non-U.S. Gov't X-ray spectrometers bent analyser crystals position-sensitive detectors |
Zusammenfassung: | Wavelength-dispersive high-resolution X-ray spectrometers often employ elastically bent crystals for the wavelength analysis. In a preceding paper [Honkanen et al. (2014). J. Synchrotron Rad. 21, 104-110] a theory for quantifying the internal stress of a macroscopically large spherically curved analyser crystal was presented. Here the theory is applied to compensate for the corresponding decrease of the energy resolution. The technique is demonstrated with a Johann-type spectrometer using a spherically bent Si(660) analyser in near-backscattering geometry, where an improvement in the energy resolution from 1.0 eV down to 0.5 eV at 9.7 keV incident photon energy was observed |
---|---|
Beschreibung: | Date Completed 30.03.2015 Date Revised 21.10.2021 published: Print-Electronic Citation Status PubMed-not-MEDLINE |
ISSN: | 1600-5775 |
DOI: | 10.1107/S1600577514011163 |