Respiratory-gated KES imaging of a rat model of acute lung injury at the Canadian Light Source

In this study, contrast-enhanced X-ray tomographic imaging for monitoring and quantifying respiratory disease in preclinical rodent models is proposed. A K-edge imaging method has been developed at the Canadian Light Source to very accurately obtain measurements of the concentration of iodinated con...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 24(2017), Pt 3 vom: 01. Mai, Seite 679-685
1. Verfasser: Deman, P (VerfasserIn)
Weitere Verfasser: Tan, S, Belev, G, Samadi, N, Martinson, M, Chapman, D, Ford, N L
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2017
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article Research Support, Non-U.S. Gov't K-edge subtraction contrast enhanced in vivo imaging lung injury respiratory gated Xenon 3H3U766W84
LEADER 01000naa a22002652 4500
001 NLM271387009
003 DE-627
005 20231224232608.0
007 cr uuu---uuuuu
008 231224s2017 xx |||||o 00| ||eng c
024 7 |a 10.1107/S160057751700193X  |2 doi 
028 5 2 |a pubmed24n0904.xml 
035 |a (DE-627)NLM271387009 
035 |a (NLM)28452761 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Deman, P  |e verfasserin  |4 aut 
245 1 0 |a Respiratory-gated KES imaging of a rat model of acute lung injury at the Canadian Light Source 
264 1 |c 2017 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 14.12.2017 
500 |a Date Revised 12.11.2023 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a In this study, contrast-enhanced X-ray tomographic imaging for monitoring and quantifying respiratory disease in preclinical rodent models is proposed. A K-edge imaging method has been developed at the Canadian Light Source to very accurately obtain measurements of the concentration of iodinated contrast agent in the pulmonary vasculature and inhaled xenon in the airspaces of rats. To compare the iodine and xenon concentration maps, a scout projection image was acquired to define the region of interest within the thorax for imaging and to ensure the same locations were imaged in each K-edge subtraction (KES) acquisition. A method for triggering image acquisition based on the real-time measurements of respiration was also developed to obtain images during end expiration when the lungs are stationary, in contrast to other previously published studies that alter the respiration to accommodate the image acquisition. In this study, images were obtained in mechanically ventilated animals using physiological parameters at the iodine K-edge in vivo and at the xenon K-edge post mortem (but still under mechanical ventilation). The imaging techniques were performed in healthy Brown Norway rats and in age-matched littermates that had an induced lung injury to demonstrate feasibility of the imaging procedures and the ability to correlate the lung injury and the quantitative measurements of contrast agent concentrations between the two KES images. The respiratory-gated KES imaging protocol can be easily adapted to image during any respiratory phase and is feasible for imaging disease models with compromised lung function 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
650 4 |a K-edge subtraction 
650 4 |a contrast enhanced 
650 4 |a in vivo imaging 
650 4 |a lung injury 
650 4 |a respiratory gated 
650 7 |a Xenon  |2 NLM 
650 7 |a 3H3U766W84  |2 NLM 
700 1 |a Tan, S  |e verfasserin  |4 aut 
700 1 |a Belev, G  |e verfasserin  |4 aut 
700 1 |a Samadi, N  |e verfasserin  |4 aut 
700 1 |a Martinson, M  |e verfasserin  |4 aut 
700 1 |a Chapman, D  |e verfasserin  |4 aut 
700 1 |a Ford, N L  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 24(2017), Pt 3 vom: 01. Mai, Seite 679-685  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:24  |g year:2017  |g number:Pt 3  |g day:01  |g month:05  |g pages:679-685 
856 4 0 |u http://dx.doi.org/10.1107/S160057751700193X  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 24  |j 2017  |e Pt 3  |b 01  |c 05  |h 679-685