Studying solutions at high shear rates : a dedicated microfluidics setup
The development of a dedicated small-angle X-ray scattering setup for the investigation of complex fluids at different controlled shear conditions is reported. The setup utilizes a microfluidics chip with a narrowing channel. As a consequence, a shear gradient is generated within the channel and the...
Veröffentlicht in: | Journal of synchrotron radiation. - 1994. - 23(2016), 2 vom: 01. März, Seite 480-6 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , , , |
Format: | Online-Aufsatz |
Sprache: | English |
Veröffentlicht: |
2016
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article Research Support, Non-U.S. Gov't aggregation beamline setup complex solutions lysozyme microfluidics protein interaction rheology shear |
Zusammenfassung: | The development of a dedicated small-angle X-ray scattering setup for the investigation of complex fluids at different controlled shear conditions is reported. The setup utilizes a microfluidics chip with a narrowing channel. As a consequence, a shear gradient is generated within the channel and the effect of shear rate on structure and interactions is mapped spatially. In a first experiment small-angle X-ray scattering is utilized to investigate highly concentrated protein solutions up to a shear rate of 300000 s(-1). These data demonstrate that equilibrium clusters of lysozyme are destabilized at high shear rates |
---|---|
Beschreibung: | Date Completed 13.12.2016 Date Revised 30.12.2020 published: Print-Electronic Citation Status MEDLINE |
ISSN: | 1600-5775 |
DOI: | 10.1107/S1600577515024856 |