Fisher Information Analysis of Depth-of-Interaction Estimation in Double-Sided Strip Detectors

In very-high-spatial-resolution gamma-ray imaging applications, such as preclinical PET and SPECT, estimation of 3D interaction location inside the detector crystal can be used to minimize parallax error in the imaging system. In this work, we investigate the effect of bias voltage setting on depth-...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science. - 1988. - 61(2014), 3 vom: 01. Juni, Seite 1243-1251
1. Verfasser: Salçin, Esen (VerfasserIn)
Weitere Verfasser: Barrett, Harrison H, Barber, H Bradford, Takeda, Shin'ichiro, Watanabe, Shin, Takahashi, Tadayuki, Furenlid, Lars R
Format: Aufsatz
Sprache:English
Veröffentlicht: 2014
Zugriff auf das übergeordnete Werk:IEEE transactions on nuclear science
Schlagworte:Journal Article CdTe CdZnTe DOI Fisher Information double-sided strip semiconductor detectors
LEADER 01000caa a22002652 4500
001 NLM250749866
003 DE-627
005 20250218193135.0
007 tu
008 231224s2014 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0835.xml 
035 |a (DE-627)NLM250749866 
035 |a (NLM)26160984 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Salçin, Esen  |e verfasserin  |4 aut 
245 1 0 |a Fisher Information Analysis of Depth-of-Interaction Estimation in Double-Sided Strip Detectors 
264 1 |c 2014 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Revised 20.11.2019 
500 |a published: Print 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a In very-high-spatial-resolution gamma-ray imaging applications, such as preclinical PET and SPECT, estimation of 3D interaction location inside the detector crystal can be used to minimize parallax error in the imaging system. In this work, we investigate the effect of bias voltage setting on depth-of-interaction (DOI) estimates for a semiconductor detector with a double-sided strip geometry. We first examine the statistical properties of the signals and develop expressions for likelihoods for given gamma-ray interaction positions. We use Fisher Information to quantify how well (in terms of variance) the measured signals can be used for DOI estimation with different bias-voltage settings. We performed measurements of detector response versus 3D position as a function of applied bias voltage by scanning with highly collimated synchrotron radiation at the Advanced Photon Source at Argonne National Laboratory. Experimental and theoretical results show that the optimum bias setting depends on whether or not the estimated event position will include the depth of interaction. We also found that for this detector geometry, the z-resolution changes with depth 
650 4 |a Journal Article 
650 4 |a CdTe 
650 4 |a CdZnTe 
650 4 |a DOI 
650 4 |a Fisher Information 
650 4 |a double-sided strip 
650 4 |a semiconductor detectors 
700 1 |a Barrett, Harrison H  |e verfasserin  |4 aut 
700 1 |a Barber, H Bradford  |e verfasserin  |4 aut 
700 1 |a Takeda, Shin'ichiro  |e verfasserin  |4 aut 
700 1 |a Watanabe, Shin  |e verfasserin  |4 aut 
700 1 |a Takahashi, Tadayuki  |e verfasserin  |4 aut 
700 1 |a Furenlid, Lars R  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on nuclear science  |d 1988  |g 61(2014), 3 vom: 01. Juni, Seite 1243-1251  |w (DE-627)NLM098149350  |x 0018-9499  |7 nnns 
773 1 8 |g volume:61  |g year:2014  |g number:3  |g day:01  |g month:06  |g pages:1243-1251 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 