Fisher Information Analysis of Depth-of-Interaction Estimation in Double-Sided Strip Detectors

In very-high-spatial-resolution gamma-ray imaging applications, such as preclinical PET and SPECT, estimation of 3D interaction location inside the detector crystal can be used to minimize parallax error in the imaging system. In this work, we investigate the effect of bias voltage setting on depth-...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science. - 1988. - 61(2014), 3 vom: 01. Juni, Seite 1243-1251
1. Verfasser: Salçin, Esen (VerfasserIn)
Weitere Verfasser: Barrett, Harrison H, Barber, H Bradford, Takeda, Shin'ichiro, Watanabe, Shin, Takahashi, Tadayuki, Furenlid, Lars R
Format: Aufsatz
Sprache:English
Veröffentlicht: 2014
Zugriff auf das übergeordnete Werk:IEEE transactions on nuclear science
Schlagworte:Journal Article CdTe CdZnTe DOI Fisher Information double-sided strip semiconductor detectors
Beschreibung
Zusammenfassung:In very-high-spatial-resolution gamma-ray imaging applications, such as preclinical PET and SPECT, estimation of 3D interaction location inside the detector crystal can be used to minimize parallax error in the imaging system. In this work, we investigate the effect of bias voltage setting on depth-of-interaction (DOI) estimates for a semiconductor detector with a double-sided strip geometry. We first examine the statistical properties of the signals and develop expressions for likelihoods for given gamma-ray interaction positions. We use Fisher Information to quantify how well (in terms of variance) the measured signals can be used for DOI estimation with different bias-voltage settings. We performed measurements of detector response versus 3D position as a function of applied bias voltage by scanning with highly collimated synchrotron radiation at the Advanced Photon Source at Argonne National Laboratory. Experimental and theoretical results show that the optimum bias setting depends on whether or not the estimated event position will include the depth of interaction. We also found that for this detector geometry, the z-resolution changes with depth