Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity

The traditional method for calculating acoustic pressure amplitude is to divide a hydrophone output voltage measurement by the hydrophone sensitivity at the acoustic working frequency, but this approach neglects frequency dependence of hydrophone sensitivity. Another method is to perform a complex d...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 61(2014), 1 vom: 24. Jan., Seite 62-75
1. Verfasser: Wear, Keith A (VerfasserIn)
Weitere Verfasser: Gammell, Paul M, Maruvada, Subha, Liu, Yunbo, Harris, Gerald R
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2014
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000naa a22002652 4500
001 NLM234279737
003 DE-627
005 20231224100931.0
007 cr uuu---uuuuu
008 231224s2014 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2014.6689776  |2 doi 
028 5 2 |a pubmed24n0781.xml 
035 |a (DE-627)NLM234279737 
035 |a (NLM)24402896 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Wear, Keith A  |e verfasserin  |4 aut 
245 1 0 |a Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity 
264 1 |c 2014 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 09.09.2014 
500 |a Date Revised 21.10.2021 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a The traditional method for calculating acoustic pressure amplitude is to divide a hydrophone output voltage measurement by the hydrophone sensitivity at the acoustic working frequency, but this approach neglects frequency dependence of hydrophone sensitivity. Another method is to perform a complex deconvolution between the hydrophone output waveform and the hydrophone impulse response (the inverse Fourier transform of the sensitivity). In this paper, the effects of deconvolution on measurements of peak compressional pressure (p+), peak rarefactional pressure (p_), and pulse intensity integral (PII) are studied. Time-delay spectrometry (TDS) was used to measure complex sensitivities from 1 to 40 MHz for 8 hydrophones used in medical ultrasound exposimetry. These included polyvinylidene fluoride (PVDF) spot-poled membrane, needle, capsule, and fiber-optic designs. Subsequently, the 8 hydrophones were used to measure a 4-cycle, 3 MHz pressure waveform mimicking a pulsed Doppler waveform. Acoustic parameters were measured for the 8 hydrophones using the traditional approach and deconvolution. Average measurements (across all 8 hydrophones) of acoustic parameters from deconvolved waveforms were 4.8 MPa (p+), 2.4 MPa (p_), and 0.21 mJ/cm(2) (PII). Compared with the traditional method, deconvolution reduced the coefficient of variation (ratio of standard deviation to mean across all 8 hydrophones) from 29% to 8% (p+), 39% to 13% (p_), and 58% to 10% (PII) 
650 4 |a Journal Article 
700 1 |a Gammell, Paul M  |e verfasserin  |4 aut 
700 1 |a Maruvada, Subha  |e verfasserin  |4 aut 
700 1 |a Liu, Yunbo  |e verfasserin  |4 aut 
700 1 |a Harris, Gerald R  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 61(2014), 1 vom: 24. Jan., Seite 62-75  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:61  |g year:2014  |g number:1  |g day:24  |g month:01  |g pages:62-75 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2014.6689776  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 61  |j 2014  |e 1  |b 24  |c 01  |h 62-75