Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity

The traditional method for calculating acoustic pressure amplitude is to divide a hydrophone output voltage measurement by the hydrophone sensitivity at the acoustic working frequency, but this approach neglects frequency dependence of hydrophone sensitivity. Another method is to perform a complex d...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 61(2014), 1 vom: 24. Jan., Seite 62-75
1. Verfasser: Wear, Keith A (VerfasserIn)
Weitere Verfasser: Gammell, Paul M, Maruvada, Subha, Liu, Yunbo, Harris, Gerald R
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2014
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article