Development and characterization of a tissue-mimicking material for high-intensity focused ultrasound

A tissue-mimicking material (TMM) for the acoustic and thermal characterization of high-intensity focused ultrasound (HIFU) devices has been developed. The material is a high-temperature hydrogel matrix (gellan gum) combined with different sizes of aluminum oxide particles and other chemicals. The u...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 58(2011), 7 vom: 15. Juli, Seite 1397-405
1. Verfasser: King, Randy L (VerfasserIn)
Weitere Verfasser: Liu, Yunbo, Maruvada, Subha, Herman, Bruce A, Wear, Keith A, Harris, Gerald R
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2011
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, U.S. Gov't, Non-P.H.S. Polysaccharides, Bacterial gellan gum 7593U09I4D Aluminum Oxide LMI26O6933
LEADER 01000naa a22002652 4500
001 NLM210038616
003 DE-627
005 20231224010904.0
007 cr uuu---uuuuu
008 231224s2011 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2011.1959  |2 doi 
028 5 2 |a pubmed24n0700.xml 
035 |a (DE-627)NLM210038616 
035 |a (NLM)21768024 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a King, Randy L  |e verfasserin  |4 aut 
245 1 0 |a Development and characterization of a tissue-mimicking material for high-intensity focused ultrasound 
264 1 |c 2011 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 10.11.2011 
500 |a Date Revised 25.11.2016 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a A tissue-mimicking material (TMM) for the acoustic and thermal characterization of high-intensity focused ultrasound (HIFU) devices has been developed. The material is a high-temperature hydrogel matrix (gellan gum) combined with different sizes of aluminum oxide particles and other chemicals. The ultrasonic properties (attenuation coefficient, speed of sound, acoustical impedance, and the thermal conductivity and diffusivity) were characterized as a function of temperature from 20 to 70°C. The backscatter coefficient and nonlinearity parameter B/A were measured at room temperature. Importantly, the attenuation coefficient has essentially linear frequency dependence, as is the case for most mammalian tissues at 37°C. The mean value is 0.64f(0.95) dB·cm(-1) at 20°C, based on measurements from 2 to 8 MHz. Most of the other relevant physical parameters are also close to the reported values, although backscatter signals are low compared with typical human soft tissues. Repeatable and consistent temperature elevations of 40°C were produced under 20-s HIFU exposures in the TMM. This TMM is appropriate for developing standardized dosimetry techniques, validating numerical models, and determining the safety and efficacy of HIFU devices 
650 4 |a Journal Article 
650 4 |a Research Support, U.S. Gov't, Non-P.H.S. 
650 7 |a Polysaccharides, Bacterial  |2 NLM 
650 7 |a gellan gum  |2 NLM 
650 7 |a 7593U09I4D  |2 NLM 
650 7 |a Aluminum Oxide  |2 NLM 
650 7 |a LMI26O6933  |2 NLM 
700 1 |a Liu, Yunbo  |e verfasserin  |4 aut 
700 1 |a Maruvada, Subha  |e verfasserin  |4 aut 
700 1 |a Herman, Bruce A  |e verfasserin  |4 aut 
700 1 |a Wear, Keith A  |e verfasserin  |4 aut 
700 1 |a Harris, Gerald R  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 58(2011), 7 vom: 15. Juli, Seite 1397-405  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:58  |g year:2011  |g number:7  |g day:15  |g month:07  |g pages:1397-405 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2011.1959  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 