Verification of the Westervelt equation for focused transducers

This study investigates the validity of the Westervelt equation for focused transducers. The angular spectrum method is employed to analyze the second-harmonic acoustic field under the weakly nonlinear approximation. Although it is well known that the Westervelt equation is accurate for the case of...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 58(2011), 5 vom: 01. Mai, Seite 1097-101
1. Verfasser: Jing, Yun (VerfasserIn)
Weitere Verfasser: Shen, Du, Clement, Greg T
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2011
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Letter
Beschreibung
Zusammenfassung:This study investigates the validity of the Westervelt equation for focused transducers. The angular spectrum method is employed to analyze the second-harmonic acoustic field under the weakly nonlinear approximation. Although it is well known that the Westervelt equation is accurate for the case of quasi-plane waves, the present work demonstrates accurate solution for the highly focused case of a spherically-curved ultrasound transducer having an aperture angle of 80°. It is further found that the solution error is inversely dependent on the nonlinearity coefficient
Beschreibung:Date Completed 08.09.2011
Date Revised 30.05.2011
published: Print
Citation Status PubMed-not-MEDLINE
ISSN:1525-8955
DOI:10.1109/TUFFC.2011.1910