Measurement uncertainty assessment of the scanning laser acoustic microscope and application to canine skin and wound

Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +/-12% and +/-15%, res...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 35(1988), 6 vom: 15., Seite 741-8
1. Verfasser: Steiger, D L (VerfasserIn)
Weitere Verfasser: O'Brien, W R, Olerud, J E, Riederer-Henderson, M A, Odland, G F
Format: Aufsatz
Sprache:English
Veröffentlicht: 1988
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM177765755
003 DE-627
005 20250209061427.0
007 tu
008 231223s1988 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0593.xml 
035 |a (DE-627)NLM177765755 
035 |a (NLM)18290211 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Steiger, D L  |e verfasserin  |4 aut 
245 1 0 |a Measurement uncertainty assessment of the scanning laser acoustic microscope and application to canine skin and wound 
264 1 |c 1988 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 02.10.2012 
500 |a Date Revised 21.02.2008 
500 |a published: Print 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +/-12% and +/-15%, respectively. Using Dow Corning 710, a silicone oil, the speed accuracy and precision are +/-2.9% and +/-0.4%, respectively. Precision was assessed using duplicate samples of canine skin and wound tissue. The estimated precisions in the measurement of the attenuation coefficient and speed were +/-16% and +/-1.7%, respectively 
650 4 |a Journal Article 
700 1 |a O'Brien, W R  |e verfasserin  |4 aut 
700 1 |a Olerud, J E  |e verfasserin  |4 aut 
700 1 |a Riederer-Henderson, M A  |e verfasserin  |4 aut 
700 1 |a Odland, G F  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 35(1988), 6 vom: 15., Seite 741-8  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:35  |g year:1988  |g number:6  |g day:15  |g pages:741-8 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 35  |j 1988  |e 6  |b 15  |h 741-8