Measurement uncertainty assessment of the scanning laser acoustic microscope and application to canine skin and wound

Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +/-12% and +/-15%, res...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 35(1988), 6 vom: 15., Seite 741-8
1. Verfasser: Steiger, D L (VerfasserIn)
Weitere Verfasser: O'Brien, W R, Olerud, J E, Riederer-Henderson, M A, Odland, G F
Format: Aufsatz
Sprache:English
Veröffentlicht: 1988
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +/-12% and +/-15%, respectively. Using Dow Corning 710, a silicone oil, the speed accuracy and precision are +/-2.9% and +/-0.4%, respectively. Precision was assessed using duplicate samples of canine skin and wound tissue. The estimated precisions in the measurement of the attenuation coefficient and speed were +/-16% and +/-1.7%, respectively
Beschreibung:Date Completed 02.10.2012
Date Revised 21.02.2008
published: Print
Citation Status PubMed-not-MEDLINE
ISSN:1525-8955