Building three-dimensional images using a time-reversal chaotic cavity

The design of two-dimensional (2-D) arrays for three-dimensional (3-D) ultrasonic imaging is a major challenge in medical and nondestructive applications. Thousands of transducers are typically needed for focusing and steering in a 3-D volume. In this article, we propose a different concept allowing...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1999. - 52(2005), 9 vom: 29. Sept., Seite 1489-97
1. Verfasser: Montaldo, Gabriel (VerfasserIn)
Weitere Verfasser: Palacio, Delphine, Tanter, Mickael, Fink, Mathias
Format: Aufsatz
Sprache:English
Veröffentlicht: 2005
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM15891161X
003 DE-627
005 20250206204818.0
007 tu
008 231223s2005 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0530.xml 
035 |a (DE-627)NLM15891161X 
035 |a (NLM)16285447 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Montaldo, Gabriel  |e verfasserin  |4 aut 
245 1 0 |a Building three-dimensional images using a time-reversal chaotic cavity 
264 1 |c 2005 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 07.12.2005 
500 |a Date Revised 17.09.2019 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a The design of two-dimensional (2-D) arrays for three-dimensional (3-D) ultrasonic imaging is a major challenge in medical and nondestructive applications. Thousands of transducers are typically needed for focusing and steering in a 3-D volume. In this article, we propose a different concept allowing us to obtain electronic 3-D focusing with a small number of transducers. The basic idea is to couple a small number of transducers to a chaotic reverberating cavity with one face in contact with the body of the patient. The reverberations of the ultrasonic waves inside the cavity create at each reflection virtual transducers. The cavity acts as an ultrasonic kaleidoscope multiplying the small number of transducers and creating a much larger virtual transducer array. By exploiting time-reversal processing, it is possible to use collectively all the virtual transducers to focus a pulse everywhere in a 3-D volume. The reception process is based on a nonlinear pulse-inversion technique in order to ensure a good contrast. The feasibility of this concept for the building of 3-D images was demonstrated using a prototype relying only on 31 emission transducers and a single reception transducer 
650 4 |a Journal Article 
700 1 |a Palacio, Delphine  |e verfasserin  |4 aut 
700 1 |a Tanter, Mickael  |e verfasserin  |4 aut 
700 1 |a Fink, Mathias  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1999  |g 52(2005), 9 vom: 29. Sept., Seite 1489-97  |w (DE-627)NLM098181017  |x 0885-3010  |7 nnns 