Building three-dimensional images using a time-reversal chaotic cavity

The design of two-dimensional (2-D) arrays for three-dimensional (3-D) ultrasonic imaging is a major challenge in medical and nondestructive applications. Thousands of transducers are typically needed for focusing and steering in a 3-D volume. In this article, we propose a different concept allowing...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1999. - 52(2005), 9 vom: 29. Sept., Seite 1489-97
1. Verfasser: Montaldo, Gabriel (VerfasserIn)
Weitere Verfasser: Palacio, Delphine, Tanter, Mickael, Fink, Mathias
Format: Aufsatz
Sprache:English
Veröffentlicht: 2005
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:The design of two-dimensional (2-D) arrays for three-dimensional (3-D) ultrasonic imaging is a major challenge in medical and nondestructive applications. Thousands of transducers are typically needed for focusing and steering in a 3-D volume. In this article, we propose a different concept allowing us to obtain electronic 3-D focusing with a small number of transducers. The basic idea is to couple a small number of transducers to a chaotic reverberating cavity with one face in contact with the body of the patient. The reverberations of the ultrasonic waves inside the cavity create at each reflection virtual transducers. The cavity acts as an ultrasonic kaleidoscope multiplying the small number of transducers and creating a much larger virtual transducer array. By exploiting time-reversal processing, it is possible to use collectively all the virtual transducers to focus a pulse everywhere in a 3-D volume. The reception process is based on a nonlinear pulse-inversion technique in order to ensure a good contrast. The feasibility of this concept for the building of 3-D images was demonstrated using a prototype relying only on 31 emission transducers and a single reception transducer
Beschreibung:Date Completed 07.12.2005
Date Revised 17.09.2019
published: Print
Citation Status MEDLINE
ISSN:0885-3010