Fast multigrid fluorescent ion chamber with 0.1 ms time response
A fast multigrid ion chamber for the detection of fluorescent X-rays has been developed. The structure of 17 grids with close separation was employed to maximize the time response as well as to give sufficient detection efficiency. The measured rise/fall response time to cyclic X-rays was shorter th...
Veröffentlicht in: | Journal of synchrotron radiation. - 1999. - 9(2002), Pt 2 vom: 01. März, Seite 99-102 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
2002
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | A fast multigrid ion chamber for the detection of fluorescent X-rays has been developed. The structure of 17 grids with close separation was employed to maximize the time response as well as to give sufficient detection efficiency. The measured rise/fall response time to cyclic X-rays was shorter than that of an existing three-grid ion chamber by more than one order of magnitude. A 0.13 ms time response was obtained at the 500 V applied voltage, where the detector can stably operate without any discharge. The available frequency range is as high as 1 kHz with a practical amplitude response |
---|---|
Beschreibung: | Date Completed 31.05.2002 Date Revised 05.06.2019 published: Print-Electronic Citation Status PubMed-not-MEDLINE |
ISSN: | 0909-0495 |