The XMaS beamline at ESRF : instrumental developments and high-resolution diffraction studies
The beamline, which is situated on a bending magnet at ESRF, comprises a unique combination of instrumentation for high-resolution and magnetic single-crystal diffraction. White-beam operation is possible, as well as focused and unfocused monochromatic modes. In addition to an eleven-axis Huber diff...
Veröffentlicht in: | Journal of synchrotron radiation. - 1999. - 8(2001), Pt 6 vom: 01. Nov., Seite 1172-81 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , , , , , , , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
2001
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | The beamline, which is situated on a bending magnet at ESRF, comprises a unique combination of instrumentation for high-resolution and magnetic single-crystal diffraction. White-beam operation is possible, as well as focused and unfocused monochromatic modes. In addition to an eleven-axis Huber diffractometer, which facilitates simple operation in both vertical and horizontal scattering geometries, there is an in-vacuum polarization analyser and slit system, mirrors for harmonic rejection, sub 4.2 K and 1 Tesla magnetic field sample environment, plus a diamond phase plate for polarization conditioning. The instrumentation developed specifically for this beamline is described, and its use illustrated by recent scientific results |
---|---|
Beschreibung: | Date Completed 14.01.2002 Date Revised 05.06.2019 published: Print-Electronic Citation Status PubMed-not-MEDLINE |
ISSN: | 0909-0495 |