The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria

• Accuracy of blood glucose monitor systems (BGMS) affects the health of patients. • All BGMS met the accuracy criteria of ISO 15197:2003. • Only three brands (six BGMS) satisfied the accuracy criteria of ISO 15197:2013. • The study further proves the necessity of implementing the new standards....

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients. - 2016. - Amsterdam [u.a.]
1. Verfasser: Chen, Huizhen (VerfasserIn)
Weitere Verfasser: Yao, Qingtao (BerichterstatterIn), Dong, Yang (BerichterstatterIn), Tang, Zhimei (BerichterstatterIn), Li, Ruiying (BerichterstatterIn), Cai, Baochao (BerichterstatterIn), Wang, Ruili (BerichterstatterIn), Chen, Qiu (BerichterstatterIn)
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2019
Zugriff auf das übergeordnete Werk:A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients
Schlagworte:Blood glucose Blood glucose monitoring system Diabetes mellitus
Umfang:7
Beschreibung
Zusammenfassung:• Accuracy of blood glucose monitor systems (BGMS) affects the health of patients. • All BGMS met the accuracy criteria of ISO 15197:2003. • Only three brands (six BGMS) satisfied the accuracy criteria of ISO 15197:2013. • The study further proves the necessity of implementing the new standards.
Beschreibung:7
DOI:10.1016/j.pcd.2018.12.010