The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria
• Accuracy of blood glucose monitor systems (BGMS) affects the health of patients. • All BGMS met the accuracy criteria of ISO 15197:2003. • Only three brands (six BGMS) satisfied the accuracy criteria of ISO 15197:2013. • The study further proves the necessity of implementing the new standards....
Veröffentlicht in: | A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients. - 2016. - Amsterdam [u.a.] |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , , , |
Format: | Online-Aufsatz |
Sprache: | English |
Veröffentlicht: |
2019
|
Zugriff auf das übergeordnete Werk: | A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients |
Schlagworte: | Blood glucose Blood glucose monitoring system Diabetes mellitus |
Umfang: | 7 |
Online verfügbar |
Volltext |