Controlling cantilevered adaptive X-ray mirrors

open access.

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 31(2024), Pt 5 vom: 01. Sept., Seite 1161-1167
1. Verfasser: Goldberg, Kenneth A (VerfasserIn)
Weitere Verfasser: La Fleche, Kyle T
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2024
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article X-ray optics adaptive optics mirrors mounting wavefront control
LEADER 01000caa a22002652 4500
001 NLM37589781X
003 DE-627
005 20240905232959.0
007 cr uuu---uuuuu
008 240806s2024 xx |||||o 00| ||eng c
024 7 |a 10.1107/S1600577524006234  |2 doi 
028 5 2 |a pubmed24n1524.xml 
035 |a (DE-627)NLM37589781X 
035 |a (NLM)39102363 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Goldberg, Kenneth A  |e verfasserin  |4 aut 
245 1 0 |a Controlling cantilevered adaptive X-ray mirrors 
264 1 |c 2024 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 05.09.2024 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a open access. 
520 |a Modeling the behavior of a prototype cantilevered X-ray adaptive mirror (held from one end) demonstrates its potential for use on high-performance X-ray beamlines. Similar adaptive mirrors are used on X-ray beamlines to compensate optical aberrations, control wavefronts and tune mirror focal distances at will. Controlled by 1D arrays of piezoceramic actuators, these glancing-incidence mirrors can provide nanometre-scale surface shape adjustment capabilities. However, significant engineering challenges remain for mounting them with low distortion and low environmental sensitivity. Finite-element analysis is used to predict the micron-scale full actuation surface shape from each channel and then linear modeling is applied to investigate the mirrors' ability to reach target profiles. Using either uniform or arbitrary spatial weighting, actuator voltages are optimized using a Moore-Penrose matrix inverse, or pseudoinverse, revealing a spatial dependence on the shape fitting with increasing fidelity farther from the mount 
650 4 |a Journal Article 
650 4 |a X-ray optics 
650 4 |a adaptive optics 
650 4 |a mirrors 
650 4 |a mounting 
650 4 |a wavefront control 
700 1 |a La Fleche, Kyle T  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 31(2024), Pt 5 vom: 01. Sept., Seite 1161-1167  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:31  |g year:2024  |g number:Pt 5  |g day:01  |g month:09  |g pages:1161-1167 
856 4 0 |u http://dx.doi.org/10.1107/S1600577524006234  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 31  |j 2024  |e Pt 5  |b 01  |c 09  |h 1161-1167