Self-Test of Air-Coupled Probe for Sensitivity Map Production Using Parabolic Reflector

The spatial distribution of a transduction efficiency over the air-coupled probe surface was proposed as a convenient tool for the probe integrity inspection. In this research, a parabolic reflector is used for passive focusing of the acoustic wave on the surface of inspected probe. Therefore, no ad...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 71(2024), 9 vom: 10. Sept., Seite 1132-1140
1. Verfasser: Svilainis, Linas (VerfasserIn)
Weitere Verfasser: Camacho Sosa Dias, Jorge, Kybartas, Darius, Chaziachmetovas, Andrius, Eidukynas, Valdas
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2024
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM374961301
003 DE-627
005 20240906232919.0
007 cr uuu---uuuuu
008 240716s2024 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2024.3428326  |2 doi 
028 5 2 |a pubmed24n1525.xml 
035 |a (DE-627)NLM374961301 
035 |a (NLM)39008401 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Svilainis, Linas  |e verfasserin  |4 aut 
245 1 0 |a Self-Test of Air-Coupled Probe for Sensitivity Map Production Using Parabolic Reflector 
264 1 |c 2024 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 05.09.2024 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a The spatial distribution of a transduction efficiency over the air-coupled probe surface was proposed as a convenient tool for the probe integrity inspection. In this research, a parabolic reflector is used for passive focusing of the acoustic wave on the surface of inspected probe. Therefore, no additional transducer is required for inspection: probe is used for self-inspection. This approach allows to avoid the expensive focused transducer and replaces it by the same-type air-coupled transducers as probe under test. Moreover, the use of the parabolic mirror for focusing is frequency-independent; therefore, such approach allows to inspect a wide range of the probes. Spread spectrum signals were used for excitation to improve the SNR and bandwidth coverage. The results of the experimental measurements of air-coupled transducer sensitivity map with natural and artificial defects implemented have been presented. Comparison with previously proposed techniques is given. It was found that defects presence distorts the focused beam, creating large sidelobes. Therefore, sensitivity map obtained with the proposed technique is lower quality than with previously proposed techniques. Beam profile measurements using a miniature microphone have been presented. Aperture-limiting mask has been proposed to reduce the sidelobes arising due to defects presence and resulting measurements quality has been improved 
650 4 |a Journal Article 
700 1 |a Camacho Sosa Dias, Jorge  |e verfasserin  |4 aut 
700 1 |a Kybartas, Darius  |e verfasserin  |4 aut 
700 1 |a Chaziachmetovas, Andrius  |e verfasserin  |4 aut 
700 1 |a Eidukynas, Valdas  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 71(2024), 9 vom: 10. Sept., Seite 1132-1140  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:71  |g year:2024  |g number:9  |g day:10  |g month:09  |g pages:1132-1140 