Evaluation of Freestyle Optium Neo H and TaiDoc-TD4235 point-of-care meters for measuring blood β-hydroxybutyrate concentration in sheep

© 2023 American Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 52(2023), 2 vom: 01. Juni, Seite 288-294
1. Verfasser: Tümer, Kenan C (VerfasserIn)
Weitere Verfasser: Kılınç, Mehmet A
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2023
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Journal Article bias ewe point-of-care device pregnancy toxemia sensitivity specificity 3-Hydroxybutyric Acid TZP1275679 Blood Glucose
LEADER 01000caa a22002652c 4500
001 NLM353655805
003 DE-627
005 20250304115556.0
007 cr uuu---uuuuu
008 231226s2023 xx |||||o 00| ||eng c
024 7 |a 10.1111/vcp.13223  |2 doi 
028 5 2 |a pubmed25n1178.xml 
035 |a (DE-627)NLM353655805 
035 |a (NLM)36859602 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Tümer, Kenan C  |e verfasserin  |4 aut 
245 1 0 |a Evaluation of Freestyle Optium Neo H and TaiDoc-TD4235 point-of-care meters for measuring blood β-hydroxybutyrate concentration in sheep 
264 1 |c 2023 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 06.06.2023 
500 |a Date Revised 06.06.2023 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a © 2023 American Society for Veterinary Clinical Pathology. 
520 |a BACKGROUND: Accurate and rapid measurement of blood β-hydroxybutyrate (β-OHB) concentrations is critical to identify hyperketonemia in sheep 
520 |a OBJECTIVE: The objective of this study was to test the performance of Freestyle Optium Neo H and TaiDoc-TD4235 point-of-care (POC) devices for measuring blood β-OHB concentrations in sheep 
520 |a METHODS: Venous blood samples from 105 sheep were evaluated with Freestyle Optium Neo H and TaiDoc-TD4235 meters. A reference approach was the laboratory measurement of serum BHB concentrations using the Randox D-3 Hydroxybutyrate reagent kit. Data were analyzed using Passing- Bablok regression and Bland-Altman analysis. The diagnostic accuracy was assessed using serum β-OHB concentration at a cut-off value of 0.8 mmol/L 
520 |a RESULTS: Passing-Bablok regression analysis revealed an intercept of 0.066 (CI95 : 0.018-0.125) and slope of 1.242 (CI95 : 1.136-1.342) for Freestyle Optium Neo H. Intercept and slope values for the TaiDoc-TD4235 meter were 0.625 (CI95 : 0.539-0.724) and 1.265 (CI95 : 1.044-1.497), respectively. The Bland-Altman plot revealed a bias of 0.224 mmol/L and 0.737 mmol/L for Freestyle Optium Neo H and TaiDoc-TD4235 meters, respectively, compared with the reference method. Observed Total Error (TEobs ) was 56.7% for the Freestyle Optium Neo H and 168.7% for the Taidoc-TD4235. The sensitivity and specificity for the Freestyle Optium Neo H were 100% and 80.9%, respectively, at the threshold of 0.8 mmol/L. The Taidoc-TD4235 demonstrated a sensitivity of 93.3% and specificity of 8.9% at the threshold of 0.8 mmol/L 
520 |a CONCLUSIONS: Because of the large TEobs , the Freestyle Optium Neo H and TaiDoc-TD4235 meters should not be used to quantify blood β-OHB concentrations in sheep 
650 4 |a Journal Article 
650 4 |a bias 
650 4 |a ewe 
650 4 |a point-of-care device 
650 4 |a pregnancy toxemia 
650 4 |a sensitivity 
650 4 |a specificity 
650 7 |a 3-Hydroxybutyric Acid  |2 NLM 
650 7 |a TZP1275679  |2 NLM 
650 7 |a Blood Glucose  |2 NLM 
700 1 |a Kılınç, Mehmet A  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Veterinary clinical pathology  |d 1975  |g 52(2023), 2 vom: 01. Juni, Seite 288-294  |w (DE-627)NLM098159984  |x 1939-165X  |7 nnas 
773 1 8 |g volume:52  |g year:2023  |g number:2  |g day:01  |g month:06  |g pages:288-294 
856 4 0 |u http://dx.doi.org/10.1111/vcp.13223  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 52  |j 2023  |e 2  |b 01  |c 06  |h 288-294