Implementing and evaluating far-field 3D X-ray diffraction at the I12 JEEP beamline, Diamond Light Source

open access.

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 29(2022), Pt 4 vom: 01. Juli, Seite 1043-1053
1. Verfasser: Ball, James A D (VerfasserIn)
Weitere Verfasser: Kareer, Anna, Magdysyuk, Oxana V, Michalik, Stefan, Vrettou, Anastasia, Parkes, Neal, Connolley, Thomas, Collins, David M
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2022
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article 3DXRD EBSD diffraction microstructure texture
LEADER 01000naa a22002652 4500
001 NLM343082527
003 DE-627
005 20231226015613.0
007 cr uuu---uuuuu
008 231226s2022 xx |||||o 00| ||eng c
024 7 |a 10.1107/S1600577522004088  |2 doi 
028 5 2 |a pubmed24n1143.xml 
035 |a (DE-627)NLM343082527 
035 |a (NLM)35787572 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Ball, James A D  |e verfasserin  |4 aut 
245 1 0 |a Implementing and evaluating far-field 3D X-ray diffraction at the I12 JEEP beamline, Diamond Light Source 
264 1 |c 2022 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 16.07.2022 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a open access. 
520 |a Three-dimensional X-ray diffraction (3DXRD) is shown to be feasible at the I12 Joint Engineering, Environmental and Processing (JEEP) beamline of Diamond Light Source. As a demonstration, a microstructually simple low-carbon ferritic steel was studied in a highly textured and annealed state. A processing pipeline suited to this beamline was created, using software already established in the 3DXRD user community, enabling grain centre-of-mass positions, orientations and strain tensor elements to be determined. Orientations, with texture measurements independently validated from electron backscatter diffraction (EBSD) data, possessed a ∼0.1° uncertainty, comparable with other 3DXRD instruments. The spatial resolution was limited by the far-field detector pixel size; the average of the grain centre of mass position errors was determined as ±∼80 µm. An average per-grain error of ∼1 × 10-3 for the elastic strains was also measured; this could be reduced in future experiments by improving sample preparation, geometry calibration, data collection and analysis techniques. Application of 3DXRD onto I12 shows great potential, where its implementation is highly desirable due to the flexible, open architecture of the beamline. User-owned or designed sample environments can be used, thus 3DXRD could be applied to previously unexplored scientific areas 
650 4 |a Journal Article 
650 4 |a 3DXRD 
650 4 |a EBSD 
650 4 |a diffraction 
650 4 |a microstructure 
650 4 |a texture 
700 1 |a Kareer, Anna  |e verfasserin  |4 aut 
700 1 |a Magdysyuk, Oxana V  |e verfasserin  |4 aut 
700 1 |a Michalik, Stefan  |e verfasserin  |4 aut 
700 1 |a Vrettou, Anastasia  |e verfasserin  |4 aut 
700 1 |a Parkes, Neal  |e verfasserin  |4 aut 
700 1 |a Connolley, Thomas  |e verfasserin  |4 aut 
700 1 |a Collins, David M  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 29(2022), Pt 4 vom: 01. Juli, Seite 1043-1053  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:29  |g year:2022  |g number:Pt 4  |g day:01  |g month:07  |g pages:1043-1053 
856 4 0 |u http://dx.doi.org/10.1107/S1600577522004088  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 29  |j 2022  |e Pt 4  |b 01  |c 07  |h 1043-1053