Reliability and concurrent validity of mobile health technology for patient self-monitoring in physical rehabilitation

© 2022 The Author(s).

Bibliographische Detailangaben
Veröffentlicht in:JSES international. - 2020. - 6(2022), 3 vom: 10. Mai, Seite 506-511
1. Verfasser: Pottorf, Ofra A (VerfasserIn)
Weitere Verfasser: Lee, Daniel J, Czujko, Paul-Neil
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2022
Zugriff auf das übergeordnete Werk:JSES international
Schlagworte:Journal Article COVID-19 Mobile health Rehabilitation Reliability Self-monitoring Smartphone Validity
LEADER 01000caa a22002652c 4500
001 NLM340952989
003 DE-627
005 20250303085412.0
007 cr uuu---uuuuu
008 231226s2022 xx |||||o 00| ||eng c
024 7 |a 10.1016/j.jseint.2022.02.002  |2 doi 
028 5 2 |a pubmed25n1136.xml 
035 |a (DE-627)NLM340952989 
035 |a (NLM)35572423 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Pottorf, Ofra A  |e verfasserin  |4 aut 
245 1 0 |a Reliability and concurrent validity of mobile health technology for patient self-monitoring in physical rehabilitation 
264 1 |c 2022 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 16.07.2022 
500 |a published: Electronic-eCollection 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a © 2022 The Author(s). 
520 |a Background: Forearm pronation and supination are important for everyday functional tasks and some recreational activities. Healthcare providers use reliable and valid tools during the physical rehabilitation process to measure joint range of motion (ROM), assess functional mobility, guide decisions for skilled interventions, and progress a plan of care. Since the onset of the COVID-19 pandemic, both healthcare providers and patients benefited from mobile health technologies that have emerged, which can be used by patients in the home to monitor ROM and assist the healthcare provider in guiding the rehabilitation process when utilizing telehealth 
520 |a Purpose: The goal of this study was to investigate the reliability and concurrent validity of a smartphone application for obtaining goniometric measurements of forearm pronation and supination 
520 |a Methods: This study consisted of 83 participants that were recruited on a voluntary basis from an academic institution. An iPhone with the application Clinometer and a standard goniometer (SG) were utilized to obtain goniometric measurements of forearm pronation and supination. The intraclass correlation coefficient (ICC) was used to analyze intrarater reliability, and the Pearson correlation coefficient was used to analyze concurrent validity. Scatterplots with regression lines were created to visually display the results 
520 |a Results: The smartphone demonstrated strong correlations for both pronation and supination (r = 0.71, P < .001; r = 0.73, P < .001). This study demonstrated overall good-excellent intrarater reliability and good concurrent validity for the smartphone application with a higher test-retest reliability in the measurement of forearm pronation compared to supination 
520 |a Conclusions: This study concludes that the reliability and concurrent validity of the smartphone was consistent with the SG for assessing forearm pronation and supination. It may be of value to further investigate interrater reliability between patient and healthcare practitioner, and report on the ease of use to assess ROM with a smartphone 
650 4 |a Journal Article 
650 4 |a COVID-19 
650 4 |a Mobile health 
650 4 |a Rehabilitation 
650 4 |a Reliability 
650 4 |a Self-monitoring 
650 4 |a Smartphone 
650 4 |a Validity 
700 1 |a Lee, Daniel J  |e verfasserin  |4 aut 
700 1 |a Czujko, Paul-Neil  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t JSES international  |d 2020  |g 6(2022), 3 vom: 10. Mai, Seite 506-511  |w (DE-627)NLM307818438  |x 2666-6383  |7 nnas 
773 1 8 |g volume:6  |g year:2022  |g number:3  |g day:10  |g month:05  |g pages:506-511 
856 4 0 |u http://dx.doi.org/10.1016/j.jseint.2022.02.002  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_50 
912 |a GBV_ILN_65 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 6  |j 2022  |e 3  |b 10  |c 05  |h 506-511