Piezoelectric Micromachined Ultrasonic Transducer for Arterial Wall Dynamics Monitoring

In this article, a [Formula: see text] piezoelectric micromachined ultrasonic transducer (PMUT) array is designed and driven with one cycle of a 5-MHz sinusoid at 10 [Formula: see text] for radial artery motion tracking. The transmit and receive performance figure of merit (FOM) of an individual PMU...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 69(2022), 1 vom: 14. Jan., Seite 291-298
1. Verfasser: Jiang, Xiaoyue (VerfasserIn)
Weitere Verfasser: Perrot, Vincent, Varray, Francois, Bart, Stephen, Hartwell, Peter G
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2022
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
LEADER 01000naa a22002652 4500
001 NLM331891662
003 DE-627
005 20231225214454.0
007 cr uuu---uuuuu
008 231225s2022 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2021.3120283  |2 doi 
028 5 2 |a pubmed24n1106.xml 
035 |a (DE-627)NLM331891662 
035 |a (NLM)34648440 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Jiang, Xiaoyue  |e verfasserin  |4 aut 
245 1 0 |a Piezoelectric Micromachined Ultrasonic Transducer for Arterial Wall Dynamics Monitoring 
264 1 |c 2022 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 10.01.2022 
500 |a Date Revised 10.01.2022 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a In this article, a [Formula: see text] piezoelectric micromachined ultrasonic transducer (PMUT) array is designed and driven with one cycle of a 5-MHz sinusoid at 10 [Formula: see text] for radial artery motion tracking. The transmit and receive performance figure of merit (FOM) of an individual PMUT over operating frequency is modeled and validated using laser Doppler vibrometer (LDV) measurements. Given a fixed cross section, the FOM inversely scales with frequency. The array aperture size is selected to obtain enough pressure and received signal to measure the radial artery wall reflection at a 5-mm depth in tissue. The 2-mm acoustic beamwidth provides enough lateral resolution for radial artery wall motion tracking. Single-line ultrasonic pulse-echo measurements with high time resolution, also called M-mode ultrasound imaging, are demonstrated to reproduce a known target motion profile with a precision of around 0.5 [Formula: see text]. In vivo radial artery dynamics are measured by placing the sensor on the wrist of a volunteer. The measured diameter change waveform of the radial artery is consistent with reports in the literature and captures key arterial pulse waveform features, including systolic upstroke, systolic decline, dicrotic notch, and diastolic runoff. The system has sufficient accuracy and precision to measure both the 50 [Formula: see text] overall diameter change and the 5- [Formula: see text] diameter change due to the dicrotic notch. A heart rate of 70 beats/min is also derived. This demonstrates the great potential of custom PMUT arrays for continuous cardiovascular system monitoring 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
700 1 |a Perrot, Vincent  |e verfasserin  |4 aut 
700 1 |a Varray, Francois  |e verfasserin  |4 aut 
700 1 |a Bart, Stephen  |e verfasserin  |4 aut 
700 1 |a Hartwell, Peter G  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 69(2022), 1 vom: 14. Jan., Seite 291-298  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:69  |g year:2022  |g number:1  |g day:14  |g month:01  |g pages:291-298 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2021.3120283  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 69  |j 2022  |e 1  |b 14  |c 01  |h 291-298