20 µs-resolved high-throughput X-ray photon correlation spectroscopy on a 500k pixel detector enabled by data-management workflow

The performance of the new 52 kHz frame rate Rigaku XSPA-500k detector was characterized on beamline 8-ID-I at the Advanced Photon Source at Argonne for X-ray photon correlation spectroscopy (XPCS) applications. Due to the large data flow produced by this detector (0.2 PB of data per 24 h of continu...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 28(2021), Pt 1 vom: 01. Jan., Seite 259-265
1. Verfasser: Zhang, Qingteng (VerfasserIn)
Weitere Verfasser: Dufresne, Eric M, Nakaye, Yasukazu, Jemian, Pete R, Sakumura, Takuto, Sakuma, Yasutaka, Ferrara, Joseph D, Maj, Piotr, Hassan, Asra, Bahadur, Divya, Ramakrishnan, Subramanian, Khan, Faisal, Veseli, Sinisa, Sandy, Alec R, Schwarz, Nicholas, Narayanan, Suresh
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2021
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article X-ray photon correlation spectroscopy (XPCS) XSPA-500k data-management workflow pixel-array detector (PAD)