Design, Development, and Operation of a Low-Intensity Focused Ultrasound Pulsation (LIFUP) System for Clinical Use
Noninvasive low-intensity focused ultrasound pulsation (LIFUP) neuromodulation provides a unique approach to both investigating and treating the brain. This work describes a well-calibrated, simple-to-use ultrasound stimulation system for neuromodulation studies. It provides a single-element 650-kHz...
Publié dans: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 68(2021), 1 vom: 21. Jan., Seite 54-64 |
---|---|
Auteur principal: | |
Autres auteurs: | , , |
Format: | Article en ligne |
Langue: | English |
Publié: |
2021
|
Accès à la collection: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control |
Sujets: | Journal Article Research Support, Non-U.S. Gov't |
Résumé: | Noninvasive low-intensity focused ultrasound pulsation (LIFUP) neuromodulation provides a unique approach to both investigating and treating the brain. This work describes a well-calibrated, simple-to-use ultrasound stimulation system for neuromodulation studies. It provides a single-element 650-kHz transducer design and a straightforward control mechanism, with extensive calibration and internal electronic monitoring to prevent unwanted over or under treatment. One goal of this approach is to relieve researchers of many of the details associated with developing their own exposure equipment. A unique transducer positioning system and distinctive MRI fiducial targets simplify alignment and targeting. The system design, control software, calibration, and alignment techniques are described in detail. Examples of transducer targeting and neurostimulation using the system are provided |
---|---|
Description: | Date Completed 25.10.2021 Date Revised 25.10.2021 published: Print-Electronic Citation Status MEDLINE |