Optimizing the u411 automated urinalysis instrument for veterinary use

© 2019 American Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 49(2020), 1 vom: 01. März, Seite 106-111
1. Verfasser: Evans, Samantha J M (VerfasserIn)
Weitere Verfasser: Sharp, Julia L, Vap, Linda M
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2020
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Journal Article canine dipstick feline reflectance photometry validation Bilirubin RFM9X3LJ49
LEADER 01000naa a22002652 4500
001 NLM305562096
003 DE-627
005 20231225121442.0
007 cr uuu---uuuuu
008 231225s2020 xx |||||o 00| ||eng c
024 7 |a 10.1111/vcp.12818  |2 doi 
028 5 2 |a pubmed24n1018.xml 
035 |a (DE-627)NLM305562096 
035 |a (NLM)31958148 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Evans, Samantha J M  |e verfasserin  |4 aut 
245 1 0 |a Optimizing the u411 automated urinalysis instrument for veterinary use 
264 1 |c 2020 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 25.01.2021 
500 |a Date Revised 25.01.2021 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a © 2019 American Society for Veterinary Clinical Pathology. 
520 |a BACKGROUND: The Cobas u411 Analyzer (Roche Diagnostics) is an automated, reflectance photometry-based urinalysis instrument designed for use with Roche's CHEMSTRIP 10UA technology and human urine samples 
520 |a OBJECTIVE: We aimed to optimize and validate the Cobas u411 Analyzer for use in canine and feline urinalysis 
520 |a METHODS: Patient urine samples presenting to the Clinical Pathology Laboratory at the Colorado State University Veterinary Teaching Hospital were analyzed with the Cobas u411 and by manual readings in parallel. Initially, 223 canine and 83 feline urine samples were run using the u411 factory settings. Following comparisons with manual results, and evaluation for directional bias, adjustments to the reflectance values were made in the instrument's programming. An additional 183 canine and 95 feline samples were run using the adjusted settings. Total urine protein concentrations were measured in 48 samples and used to generate receiver operating characteristic curves for the protein test pad 
520 |a RESULTS: Following adjustments in reflectance programming, concordance between u411 and manual results was increased by 17.7% for protein, 11.7% for ketones, and 4.5% for bilirubin. Concordances for pH, glucose, and blood were not substantially changed. Discordance for all analytes was ≤3%. Canine and feline samples had similar levels of discordance, though marginal concordance was higher in dogs for ketones, bilirubin, and blood 
520 |a CONCLUSIONS: Adjustments to the reflectance programming of the Cobas u411 Analyzer improved concordance with manual results for canine and feline samples. This instrument has the potential to greatly increase both efficiency and consistency of urinalysis procedures in higher throughput veterinary diagnostic laboratories 
650 4 |a Journal Article 
650 4 |a canine 
650 4 |a dipstick 
650 4 |a feline 
650 4 |a reflectance photometry 
650 4 |a validation 
650 7 |a Bilirubin  |2 NLM 
650 7 |a RFM9X3LJ49  |2 NLM 
700 1 |a Sharp, Julia L  |e verfasserin  |4 aut 
700 1 |a Vap, Linda M  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Veterinary clinical pathology  |d 1975  |g 49(2020), 1 vom: 01. März, Seite 106-111  |w (DE-627)NLM098159984  |x 1939-165X  |7 nnns 
773 1 8 |g volume:49  |g year:2020  |g number:1  |g day:01  |g month:03  |g pages:106-111 
856 4 0 |u http://dx.doi.org/10.1111/vcp.12818  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 49  |j 2020  |e 1  |b 01  |c 03  |h 106-111