Visibility of Bioresorbable Vascular Scaffold in Intravascular Ultrasound Imaging

Bioresorbable vascular scaffold (BVS) has recently been spotlighted for its unique characteristics of absorbing into blood vessels and eventually disappearing. Although intravascular ultrasound (IVUS) is the most common guiding tool for stent deployment, the echogenicity of BVS struts has changed as...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 67(2020), 6 vom: 16. Juni, Seite 1090-1101
1. Verfasser: Choi, Taewon (VerfasserIn)
Weitere Verfasser: Yu, Hwanseung, Chang, Seoyun, Ha, Dong-Heon, Cho, Dong-Woo, Jang, Jinah, Lee, Changyang, Lu, Gengxi, Chang, Jin Ho, Zhou, Qifa, Park, Jinhyoung
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2020
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, N.I.H., Extramural Research Support, Non-U.S. Gov't
Beschreibung
Zusammenfassung:Bioresorbable vascular scaffold (BVS) has recently been spotlighted for its unique characteristics of absorbing into blood vessels and eventually disappearing. Although intravascular ultrasound (IVUS) is the most common guiding tool for stent deployment, the echogenicity of BVS struts has changed as the center of stent lumen and scanning rotation is not concentric, which may cause a critical erroneous measurement in practice. This study investigated the physical conditions for dimming the stent brightness in IVUS images using a finite-difference method (FDM) to numerically solve acoustic wave propagation through nonhomogeneous medium. The dimmed brightness is caused by an angled rectangular cross section of a strut and its similar acoustic impedance with water. Imaging frequency is not a major cause. However, the angle between the acoustic beam and the BVS surface is the major cause of the dimmed brightness. As a solution, an approach using a frequency compounding method with signal polarity comparator was proposed to recover the reduced brightness without sacrificing spatial resolutions. Based on the simulation study, the signal level from BVS can be attenuated down by 17 dB when the angle between the acoustic beamline and the surface of BVS is more than 45°. With the proposed frequency compounding approach, the reduced signal can be recovered by 6 dB. In the experimental BVS IVUS imaging, strut brightness was reduced by 18 dB with an angled strut position and recovered by 5 dB with the proposed frequency compounding method. A pig coronary was imaged to demonstrate the performance of the proposed method