Broadband Ultrasonic Array Transducer From Multilayer Piezoelectric Ceramic With Lowered Co-Firing Temperature

Increasing array transducer bandwidth (BW) and signal-to-noise ratio (SNR) is a critical issue for producing a high-quality medical ultrasound image. However, array elements with small size tend to have poor sensitivity due to a much higher impedance compared with the electrical impedance of the tra...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 67(2020), 5 vom: 01. Mai, Seite 968-974
1. Verfasser: Ke, Qingqing (VerfasserIn)
Weitere Verfasser: Liew, Weng Heng, Zhang, Lei, Tan, Chee Kiang Ivan, Jiang, Changyun, Yao, Kui
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2020
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
Beschreibung
Zusammenfassung:Increasing array transducer bandwidth (BW) and signal-to-noise ratio (SNR) is a critical issue for producing a high-quality medical ultrasound image. However, array elements with small size tend to have poor sensitivity due to a much higher impedance compared with the electrical impedance of the transmitter and receiver circuit. Implementation of multilayer ceramic (MLC) is an effective way of reducing impedance, and thus, with a potential for improving SNR for an ultrasonic probe. In this work, we fabricated multilayer piezoelectric ceramic with a composition of 0.1Pb(Ni1/3Nb2/3)O3-0.35Pb(Zn1/3Nb2/3)O3-0.15Pb(Mg1/3Nb2/3)O3-0.1PbZrO3-0.3PbTiO3-4mol% excess NiO (PNN-PZN-PMN-PZ-PT), by a roll to roll tape casting process and co-fired with 90Ag/10Pd electrode at a low temperature of 950 °C. Using five-layer MLC (5L-MLC) as obtained, we designed and demonstrated a 5 MHz 32-element array transducer for ultrasonic and photoacoustic imaging. The five-layer transducer element exhibited a BW of 87% at -6 dB, substantially higher than 62% for single-layer ceramic (SLC) element. In addition, the insertion loss was improved by 16.2 dB over the SLC element with an external impedance of 50 Ω . Both the experimental results and theoretical analysis showed that our array transducer made of the PNN-PZN-PMN-PZ-PT MLC is promising for acquiring high-quality ultrasonic and photoacoustic images
Beschreibung:Date Completed 22.02.2021
Date Revised 22.02.2021
published: Print-Electronic
Citation Status MEDLINE
ISSN:1525-8955
DOI:10.1109/TUFFC.2019.2959038