Upgrade of laser pump time-resolved X-ray probes in Beijing synchrotron

The upgrade of the laser pump time-resolved X-ray probes, namely time-resolved X-ray absorption spectroscopy (TR-XAS) and X-ray diffraction (TR-XRD), implemented at the Beijing Synchrotron Radiation Facility, is described. The improvements include a superbunch fill, a high-efficiency fluorescence co...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 26(2019), Pt 6 vom: 01. Nov., Seite 2075-2080
1. Verfasser: Yu, Can (VerfasserIn)
Weitere Verfasser: Zhang, Yan, Lei, Ge, Wang, Hao, Wang, Yu Hang, Zhang, Bing Bing, Sun, Da Rui, Zhao, Jia Li, Zhou, Yang Fan, Li, Qiu Ju, Gao, Zhen Hua, Zhan, Fei, Tao, Ye
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2019
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article laser pump/X-ray probe thin films time-resolved X-ray absorption spectroscopy time-resolved X-ray diffraction
Beschreibung
Zusammenfassung:The upgrade of the laser pump time-resolved X-ray probes, namely time-resolved X-ray absorption spectroscopy (TR-XAS) and X-ray diffraction (TR-XRD), implemented at the Beijing Synchrotron Radiation Facility, is described. The improvements include a superbunch fill, a high-efficiency fluorescence collection, an efficient spatial overlap protocol and a new data-acquisition scheme. After upgrade, the adequate TR-XAS signal is now obtained in a 0.3 mM solution, compared with a 6 mM solution in our previous report. Furthermore, to extend application in photophysics, the TR-XAS probe is applied on SrCoO2.5 thin film. And for the first time, TR-XAS is combined with TR-XRD to simultaneously detect the kinetic trace of structural changes in thin film
Beschreibung:Date Completed 18.11.2019
Date Revised 18.11.2019
published: Print-Electronic
Citation Status PubMed-not-MEDLINE
ISSN:1600-5775