Reducing RF-induced Heating near Implanted Leads through High-Dielectric Capacitive Bleeding of Current (CBLOC)

Patients with implanted medical devices such as deep brain stimulation or spinal cord stimulation are often unable to receive magnetic resonance imaging (MRI). This is because once the device is within the radiofrequency (RF) field of the MRI scanner, electrically conductive leads act as antenna, am...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on microwave theory and techniques. - 1998. - 67(2019), 3 vom: 12. März, Seite 1265-1273
1. Verfasser: Golestanirad, Laleh (VerfasserIn)
Weitere Verfasser: Angelone, Leonardo M, Kirsch, John, Downs, Sean, Keil, Boris, Bonmassar, Giorgio, Wald, Lawrence L
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2019
Zugriff auf das übergeordnete Werk:IEEE transactions on microwave theory and techniques
Schlagworte:Journal Article MR Conditional RF heating RF safety electrode leads finite element method high dielectric material high field magnetic resonance imaging (MRI) medical implants mehr... numerical simulations specific absorption rate (SAR)
LEADER 01000caa a22002652 4500
001 NLM302145346
003 DE-627
005 20240720232447.0
007 cr uuu---uuuuu
008 231225s2019 xx |||||o 00| ||eng c
024 7 |a 10.1109/TMTT.2018.2885517  |2 doi 
028 5 2 |a pubmed24n1476.xml 
035 |a (DE-627)NLM302145346 
035 |a (NLM)31607756 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Golestanirad, Laleh  |e verfasserin  |4 aut 
245 1 0 |a Reducing RF-induced Heating near Implanted Leads through High-Dielectric Capacitive Bleeding of Current (CBLOC) 
264 1 |c 2019 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 20.07.2024 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a Patients with implanted medical devices such as deep brain stimulation or spinal cord stimulation are often unable to receive magnetic resonance imaging (MRI). This is because once the device is within the radiofrequency (RF) field of the MRI scanner, electrically conductive leads act as antenna, amplifying the RF energy deposition in the tissue and causing possible excessive tissue heating. Here we propose a novel concept in lead design in which 40cm lead wires are coated with a ~1.2mm layer of high dielectric constant material (155 < ε r < 250) embedded in a weakly conductive insulation (σ = 20S/m). The technique called High-Dielectric Capacitive Bleeding of Current, or CBLOC, works by forming a distributed capacitance along the lengths of the lead, efficiently dissipating RF energy before it reaches the exposed tip. Measurements during RF exposure at 64 MHz and 123 MHz demonstrated that CBLOC leads generated 20-fold less heating at 1.5 T, and 40-fold less heating at 3 T compared to control leads. Numerical simulations of RF exposure at 297 MHz (7T) predicted a 15-fold reduction in specific absorption rate (SAR) of RF energy around the tip of CBLOC leads compared to control leads 
650 4 |a Journal Article 
650 4 |a MR Conditional 
650 4 |a RF heating 
650 4 |a RF safety 
650 4 |a electrode leads 
650 4 |a finite element method 
650 4 |a high dielectric material 
650 4 |a high field 
650 4 |a magnetic resonance imaging (MRI) 
650 4 |a medical implants 
650 4 |a numerical simulations 
650 4 |a specific absorption rate (SAR) 
700 1 |a Angelone, Leonardo M  |e verfasserin  |4 aut 
700 1 |a Kirsch, John  |e verfasserin  |4 aut 
700 1 |a Downs, Sean  |e verfasserin  |4 aut 
700 1 |a Keil, Boris  |e verfasserin  |4 aut 
700 1 |a Bonmassar, Giorgio  |e verfasserin  |4 aut 
700 1 |a Wald, Lawrence L  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on microwave theory and techniques  |d 1998  |g 67(2019), 3 vom: 12. März, Seite 1265-1273  |w (DE-627)NLM098148141  |x 0018-9480  |7 nnns 
773 1 8 |g volume:67  |g year:2019  |g number:3  |g day:12  |g month:03  |g pages:1265-1273 
856 4 0 |u http://dx.doi.org/10.1109/TMTT.2018.2885517  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 67  |j 2019  |e 3  |b 12  |c 03  |h 1265-1273