A zone-plate-based two-color spectrometer for indirect X-ray absorption spectroscopy

open access.

Détails bibliographiques
Publié dans:Journal of synchrotron radiation. - 1994. - 26(2019), Pt 4 vom: 01. Juli, Seite 1266-1271
Auteur principal: Döring, Florian (Auteur)
Autres auteurs: Risch, Marcel, Rösner, Benedikt, Beye, Martin, Busse, Philipp, Kubiček, Katharina, Glaser, Leif, Miedema, Piter S, Soltau, Jakob, Raiser, Dirk, Guzenko, Vitaliy A, Szabadics, Lukas, Kochanneck, Leif, Baumung, Max, Buck, Jens, Jooss, Christian, Techert, Simone, David, Christian
Format: Article en ligne
Langue:English
Publié: 2019
Accès à la collection:Journal of synchrotron radiation
Sujets:Journal Article soft X-rays two-color X-ray absorption spectroscopy zone plates
Description
Résumé:open access.
X-ray absorption spectroscopy (XAS) is a powerful element-specific technique that allows the study of structural and chemical properties of matter. Often an indirect method is used to access the X-ray absorption (XA). This work demonstrates a new XAS implementation that is based on off-axis transmission Fresnel zone plates to obtain the XA spectrum of La0.6Sr0.4MnO3 by analysis of three emission lines simultaneously at the detector, namely the O 2p-1s, Mn 3s-2p and Mn 3d-2p transitions. This scheme allows the simultaneous measurement of an integrated total fluorescence yield and the partial fluorescence yields (PFY) of the Mn 3s-2p and Mn 3d-2p transitions when scanning the Mn L-edge. In addition to this, the reduction in O fluorescence provides another measure for absorption often referred to as the inverse partial fluorescence yield (IPFY). Among these different methods to measure XA, the Mn 3s PFY and IPFY deviate the least from the true XA spectra due to the negligible influence of selection rules on the decay channel. Other advantages of this new scheme are the potential to strongly increase the efficiency and throughput compared with similar measurements using conventional gratings and to increase the signal-to-noise of the XA spectra as compared with a photodiode. The ability to record undistorted bulk XA spectra at high flux is crucial for future in situ spectroscopy experiments on complex materials
Description:Date Revised 25.02.2020
published: Print-Electronic
Citation Status PubMed-not-MEDLINE
ISSN:1600-5775
DOI:10.1107/S1600577519003898