Signal Coherence and Image Amplitude With the Filtered Delay Multiply and Sum Beamformer

The filtered delay multiply and sum (F-DMAS) beamformer has recently been presented in the context of medical ultrasound image formation. This nonlinear beamformer produces images with improved contrast resolution and noise rejection when compared with the delay and sum (DAS) beamformer. In an attem...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 65(2018), 7 vom: 11. Juli, Seite 1133-1140
1. Verfasser: Prieur, Fabrice (VerfasserIn)
Weitere Verfasser: Rindal, Ole Marius Hoel, Austeng, Andreas
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2018
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
LEADER 01000naa a22002652 4500
001 NLM28635912X
003 DE-627
005 20231225051527.0
007 cr uuu---uuuuu
008 231225s2018 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2018.2831789  |2 doi 
028 5 2 |a pubmed24n0954.xml 
035 |a (DE-627)NLM28635912X 
035 |a (NLM)29993368 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Prieur, Fabrice  |e verfasserin  |4 aut 
245 1 0 |a Signal Coherence and Image Amplitude With the Filtered Delay Multiply and Sum Beamformer 
264 1 |c 2018 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 03.09.2019 
500 |a Date Revised 03.09.2019 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a The filtered delay multiply and sum (F-DMAS) beamformer has recently been presented in the context of medical ultrasound image formation. This nonlinear beamformer produces images with improved contrast resolution and noise rejection when compared with the delay and sum (DAS) beamformer. In an attempt to better understand the origin of the improved image quality, this paper shows a theoretical study of the image amplitude statistics backed up by numerical simulations. The results show that the difference in image amplitude using the DAS or F-DMAS beamformers can be partly explained by the way signal coherence influences both beamformers. When using the F-DMAS compared with the DAS beamformer, the image amplitude is shown to be more dependent on the signal coherence. Experimental ultrasound images of a phantom confirm our findings 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
700 1 |a Rindal, Ole Marius Hoel  |e verfasserin  |4 aut 
700 1 |a Austeng, Andreas  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 65(2018), 7 vom: 11. Juli, Seite 1133-1140  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:65  |g year:2018  |g number:7  |g day:11  |g month:07  |g pages:1133-1140 