X-ray beam-position feedback system with easy-to-use beam-position monitor
X-ray beam-position stability is indispensable in cutting-edge experiments using synchrotron radiation. Here, for the first time, a beam-position feedback system is presented that utilizes an easy-to-use X-ray beam-position monitor incorporating a diamond-fluorescence screen. The acceptable range of...
Publié dans: | Journal of synchrotron radiation. - 1994. - 25(2018), Pt 3 vom: 01. Mai, Seite 869-873 |
---|---|
Auteur principal: | |
Autres auteurs: | , , |
Format: | Article en ligne |
Langue: | English |
Publié: |
2018
|
Accès à la collection: | Journal of synchrotron radiation |
Sujets: | Journal Article X-ray beam-position monitor diamond-fluorescence screen feedback system |
Résumé: | X-ray beam-position stability is indispensable in cutting-edge experiments using synchrotron radiation. Here, for the first time, a beam-position feedback system is presented that utilizes an easy-to-use X-ray beam-position monitor incorporating a diamond-fluorescence screen. The acceptable range of the monitor is above 500 µm and the feedback system maintains the beam position within 3 µm. In addition to being inexpensive, the system has two key advantages: it works without a scale factor for position calibration, and it has no dependence on X-ray energy, X-ray intensity, beam size or beam shape |
---|---|
Description: | Date Revised 20.11.2019 published: Print-Electronic Citation Status PubMed-not-MEDLINE |