Principles around Accurate Blood Volume Collection Using Capillary Action

Capillary action is one mechanism microfluidics uses to draw liquid autonomously in a substrate without the need of external energy. This behavior can be exploited to collect accurate volumes of liquids such as blood in narrow columns known as capillary tubes and help the development of inexpensive,...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Langmuir : the ACS journal of surfaces and colloids. - 1992. - 33(2017), 50 vom: 19. Dez., Seite 14220-14225
1. Verfasser: Lapierre, Florian (VerfasserIn)
Weitere Verfasser: Gooley, Andrew, Breadmore, Michael
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2017
Zugriff auf das übergeordnete Werk:Langmuir : the ACS journal of surfaces and colloids
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
LEADER 01000naa a22002652 4500
001 NLM278271707
003 DE-627
005 20231225020157.0
007 cr uuu---uuuuu
008 231225s2017 xx |||||o 00| ||eng c
024 7 |a 10.1021/acs.langmuir.7b02825  |2 doi 
028 5 2 |a pubmed24n0927.xml 
035 |a (DE-627)NLM278271707 
035 |a (NLM)29161045 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Lapierre, Florian  |e verfasserin  |4 aut 
245 1 0 |a Principles around Accurate Blood Volume Collection Using Capillary Action 
264 1 |c 2017 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 18.01.2019 
500 |a Date Revised 18.01.2019 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a Capillary action is one mechanism microfluidics uses to draw liquid autonomously in a substrate without the need of external energy. This behavior can be exploited to collect accurate volumes of liquids such as blood in narrow columns known as capillary tubes and help the development of inexpensive, user-friendly personalized biomedical tools. Precision bore glass capillaries demonstrate the "state of the art" for volume accuracy and precision, but height and radius must be carefully chosen in order to exploit the capillary action behavior efficiently. This Article investigates the influence of surface glass aging, due to prolonged exposure to humid air, and hematocrit level on the blood capillary rise. It provides also the tools to correctly define the optimum capillary dimensions to collect an accurate volume of blood in a glass capillary tube 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
700 1 |a Gooley, Andrew  |e verfasserin  |4 aut 
700 1 |a Breadmore, Michael  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Langmuir : the ACS journal of surfaces and colloids  |d 1992  |g 33(2017), 50 vom: 19. Dez., Seite 14220-14225  |w (DE-627)NLM098181009  |x 1520-5827  |7 nnns 
773 1 8 |g volume:33  |g year:2017  |g number:50  |g day:19  |g month:12  |g pages:14220-14225 