Non-uniform nanosecond gate-delay of hybrid pixel detectors

A simple experiment to characterize the gating properties of X-ray area detectors using pulsed X-ray sources is presented. For a number of time-resolved experiments the gating uniformity of area detectors is important. Relative gating delays between individual modules and readout chips of PILATUS2 s...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 24(2017), Pt 5 vom: 01. Sept., Seite 1082-1085
1. Verfasser: Shayduk, Roman (VerfasserIn)
Weitere Verfasser: Pennicard, David, Krausert, Konstantin, Gaal, Peter, Volkov, Sergey, Vonk, Vedran, Hejral, Uta, Jankowski, Maciej, Reinhardt, Matthias, Leitenberger, Wolfram, Stierle, Andreas
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2017
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article hybrid pixel X-ray detector jitter pump–probe X-ray diffraction timing triggering
LEADER 01000naa a22002652 4500
001 NLM275348741
003 DE-627
005 20231225005456.0
007 cr uuu---uuuuu
008 231225s2017 xx |||||o 00| ||eng c
024 7 |a 10.1107/S1600577517009158  |2 doi 
028 5 2 |a pubmed24n0917.xml 
035 |a (DE-627)NLM275348741 
035 |a (NLM)28862632 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Shayduk, Roman  |e verfasserin  |4 aut 
245 1 0 |a Non-uniform nanosecond gate-delay of hybrid pixel detectors 
264 1 |c 2017 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 20.11.2019 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a A simple experiment to characterize the gating properties of X-ray area detectors using pulsed X-ray sources is presented. For a number of time-resolved experiments the gating uniformity of area detectors is important. Relative gating delays between individual modules and readout chips of PILATUS2 series area X-ray detectors have been observed. For three modules of a PILATUS 300K-W unit the maximum gating offset between the modules is found to be as large as 30 ns. On average, the first photosensor module is found to be triggered 15 ns and 30 ns later than the second and the third modules, respectively 
650 4 |a Journal Article 
650 4 |a hybrid pixel X-ray detector 
650 4 |a jitter 
650 4 |a pump–probe X-ray diffraction 
650 4 |a timing 
650 4 |a triggering 
700 1 |a Pennicard, David  |e verfasserin  |4 aut 
700 1 |a Krausert, Konstantin  |e verfasserin  |4 aut 
700 1 |a Gaal, Peter  |e verfasserin  |4 aut 
700 1 |a Volkov, Sergey  |e verfasserin  |4 aut 
700 1 |a Vonk, Vedran  |e verfasserin  |4 aut 
700 1 |a Hejral, Uta  |e verfasserin  |4 aut 
700 1 |a Jankowski, Maciej  |e verfasserin  |4 aut 
700 1 |a Reinhardt, Matthias  |e verfasserin  |4 aut 
700 1 |a Leitenberger, Wolfram  |e verfasserin  |4 aut 
700 1 |a Stierle, Andreas  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 24(2017), Pt 5 vom: 01. Sept., Seite 1082-1085  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:24  |g year:2017  |g number:Pt 5  |g day:01  |g month:09  |g pages:1082-1085 
856 4 0 |u http://dx.doi.org/10.1107/S1600577517009158  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 24  |j 2017  |e Pt 5  |b 01  |c 09  |h 1082-1085