Development of optical choppers for time-resolved measurements at soft X-ray synchrotron radiation beamlines

Two types of optical choppers for time-resolved measurements at synchrotron radiation soft X-ray beamlines have been developed. One type uses an air-spindle-type rotation mechanism with a two-stage differential pumping system to maintain the ultra-high vacuum of the X-ray beamline, and the other use...

Description complète

Détails bibliographiques
Publié dans:Journal of synchrotron radiation. - 1994. - 24(2017), Pt 3 vom: 01. Mai, Seite 560-565
Auteur principal: Osawa, Hitoshi (Auteur)
Autres auteurs: Ohkochi, Takuo, Fujisawa, Masami, Kimura, Shigeru, Kinoshita, Toyohiko
Format: Article en ligne
Langue:English
Publié: 2017
Accès à la collection:Journal of synchrotron radiation
Sujets:Journal Article Research Support, Non-U.S. Gov't X-ray chopper soft X-ray time-resolved measurements ultra-high vacuum
LEADER 01000caa a22002652c 4500
001 NLM27138686X
003 DE-627
005 20250221135627.0
007 cr uuu---uuuuu
008 231224s2017 xx |||||o 00| ||eng c
024 7 |a 10.1107/S1600577517002399  |2 doi 
028 5 2 |a pubmed25n0904.xml 
035 |a (DE-627)NLM27138686X 
035 |a (NLM)28452746 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Osawa, Hitoshi  |e verfasserin  |4 aut 
245 1 0 |a Development of optical choppers for time-resolved measurements at soft X-ray synchrotron radiation beamlines 
264 1 |c 2017 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 11.07.2017 
500 |a Date Revised 13.11.2018 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a Two types of optical choppers for time-resolved measurements at synchrotron radiation soft X-ray beamlines have been developed. One type uses an air-spindle-type rotation mechanism with a two-stage differential pumping system to maintain the ultra-high vacuum of the X-ray beamline, and the other uses a magnetic bearing. Both can be installed at the soft X-ray beamlines at SPring-8, greatly improving the accessibility of pump-and-probe spectroscopy. The combination of X-ray chopper and pump-and-probe photoemission electron microscope at SPring-8 provides drastic improvements in signal-to-noise ratio and resolution compared with techniques using high-voltage gating of channel plate detectors. The choppers have the capability to be used not only at synchrotron radiation facilities but also at other types of soft X-ray and VUV beamlines 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
650 4 |a X-ray chopper 
650 4 |a soft X-ray 
650 4 |a time-resolved measurements 
650 4 |a ultra-high vacuum 
700 1 |a Ohkochi, Takuo  |e verfasserin  |4 aut 
700 1 |a Fujisawa, Masami  |e verfasserin  |4 aut 
700 1 |a Kimura, Shigeru  |e verfasserin  |4 aut 
700 1 |a Kinoshita, Toyohiko  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 24(2017), Pt 3 vom: 01. Mai, Seite 560-565  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnas 
773 1 8 |g volume:24  |g year:2017  |g number:Pt 3  |g day:01  |g month:05  |g pages:560-565 