High Stability Comparison of Atomic Fountains using two Different Cryogenic Oscillators

We present a detailed characterization of two atomic clock interrogation systems based on two different cryogenic sapphire oscillators operated simultaneously. We use them as references for two accurate fountain clock frequency standards participating in international atomic time and operating both...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 63(2016), 8 vom: 15. Aug., Seite 1198-1203
1. Verfasser: Abgrall, Michel (VerfasserIn)
Weitere Verfasser: Guena, Jocelyne, Lours, Michel, Santarelli, Giorgio, Tobar, Michael, Bize, Sebastien, Grop, Serge, Dubois, Benoit, Fluhr, Christophe, Giordano, Vincent
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2016
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000naa a22002652 4500
001 NLM268250138
003 DE-627
005 20231224222315.0
007 cr uuu---uuuuu
008 231224s2016 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2016.2570898  |2 doi 
028 5 2 |a pubmed24n0894.xml 
035 |a (DE-627)NLM268250138 
035 |a (NLM)28113421 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Abgrall, Michel  |e verfasserin  |4 aut 
245 1 0 |a High Stability Comparison of Atomic Fountains using two Different Cryogenic Oscillators 
264 1 |c 2016 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Revised 20.11.2019 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a We present a detailed characterization of two atomic clock interrogation systems based on two different cryogenic sapphire oscillators operated simultaneously. We use them as references for two accurate fountain clock frequency standards participating in international atomic time and operating both at the quantum projection noise frequency limit. The two fountain comparison down to a few 10 16 over 28 days demonstrates the potential of a cryocooled oscillator to replace a He refilled cryogenic oscillator 
650 4 |a Journal Article 
700 1 |a Guena, Jocelyne  |e verfasserin  |4 aut 
700 1 |a Lours, Michel  |e verfasserin  |4 aut 
700 1 |a Santarelli, Giorgio  |e verfasserin  |4 aut 
700 1 |a Tobar, Michael  |e verfasserin  |4 aut 
700 1 |a Bize, Sebastien  |e verfasserin  |4 aut 
700 1 |a Grop, Serge  |e verfasserin  |4 aut 
700 1 |a Dubois, Benoit  |e verfasserin  |4 aut 
700 1 |a Fluhr, Christophe  |e verfasserin  |4 aut 
700 1 |a Giordano, Vincent  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 63(2016), 8 vom: 15. Aug., Seite 1198-1203  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:63  |g year:2016  |g number:8  |g day:15  |g month:08  |g pages:1198-1203 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2016.2570898  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 63  |j 2016  |e 8  |b 15  |c 08  |h 1198-1203