Fabrication of Two Flow Phantoms for Doppler Ultrasound Imaging

Flow phantoms are widely used in studies associated with Doppler ultrasound measurements, acting as an effective experimental validation system in cardiovascular-related research and in new algorithm/instrumentation development. The development of materials that match the acoustic and mechanical pro...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 64(2017), 1 vom: 07. Jan., Seite 53-65
1. Verfasser: Zhou, Xiaowei (VerfasserIn)
Weitere Verfasser: Kenwright, David A, Wang, Shiying, Hossack, John A, Hoskins, Peter R
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2017
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
LEADER 01000caa a22002652 4500
001 NLM266890733
003 DE-627
005 20250221000236.0
007 cr uuu---uuuuu
008 231224s2017 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2016.2634919  |2 doi 
028 5 2 |a pubmed25n0889.xml 
035 |a (DE-627)NLM266890733 
035 |a (NLM)27925588 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Zhou, Xiaowei  |e verfasserin  |4 aut 
245 1 0 |a Fabrication of Two Flow Phantoms for Doppler Ultrasound Imaging 
264 1 |c 2017 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 26.11.2018 
500 |a Date Revised 21.03.2022 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a Flow phantoms are widely used in studies associated with Doppler ultrasound measurements, acting as an effective experimental validation system in cardiovascular-related research and in new algorithm/instrumentation development. The development of materials that match the acoustic and mechanical properties of the vascular system is of great interest while designing flow phantoms. Although recipes that meet the flow phantom standard defined by the International Electrotechnical Commission 61685 are already available in the literature, the standard procedure for material preparations and phantom fabrications has not been well established. In this paper, two types of flow phantoms, with and without blood vessel mimic, are described in detail in terms of the material preparation and phantom fabrication. The phantom materials chosen for the two phantoms are from published phantom studies, and their physical properties have been investigated previously. Both the flow phantoms have been scanned by ultrasound scanners and images from different modes are presented. These phantoms may be used in the validation and characterization of Doppler ultrasound measurements in blood vessels with a diameter above 1 mm 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
700 1 |a Kenwright, David A  |e verfasserin  |4 aut 
700 1 |a Wang, Shiying  |e verfasserin  |4 aut 
700 1 |a Hossack, John A  |e verfasserin  |4 aut 
700 1 |a Hoskins, Peter R  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 64(2017), 1 vom: 07. Jan., Seite 53-65  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:64  |g year:2017  |g number:1  |g day:07  |g month:01  |g pages:53-65 