Effect of prewarming EDTA blood samples to 37°C on platelet count measured by Sysmex XT-2000iV in dogs, cats, and horses

© 2016 American Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 45(2016), 3 vom: 16. Sept., Seite 444-9
1. Verfasser: Williams, Tim L (VerfasserIn)
Weitere Verfasser: Archer, Joy
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2016
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Journal Article Platelet clumping preanalytic error pseudothrombocytopenia Edetic Acid 9G34HU7RV0
LEADER 01000caa a22002652 4500
001 NLM262186187
003 DE-627
005 20250220091737.0
007 cr uuu---uuuuu
008 231224s2016 xx |||||o 00| ||eng c
024 7 |a 10.1111/vcp.12378  |2 doi 
028 5 2 |a pubmed25n0873.xml 
035 |a (DE-627)NLM262186187 
035 |a (NLM)27391303 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Williams, Tim L  |e verfasserin  |4 aut 
245 1 0 |a Effect of prewarming EDTA blood samples to 37°C on platelet count measured by Sysmex XT-2000iV in dogs, cats, and horses 
264 1 |c 2016 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 07.03.2018 
500 |a Date Revised 02.12.2018 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a © 2016 American Society for Veterinary Clinical Pathology. 
520 |a BACKGROUND: Pseudothrombocytopenia secondary to platelet clumping is a common cause of preanalytic error for platelet counts in dogs, cats, and horses. In human beings, it is suggested that prewarming blood samples to 37°C prior to hematology analysis will reduce platelet clumping 
520 |a OBJECTIVES: The purpose of the study was to evaluate the effect of prewarming EDTA blood samples to 37°C on measured platelet counts and other hematologic variables 
520 |a METHODS: The EDTA blood samples from dogs, cats and horses submitted to the clinical pathology laboratory at the University of Cambridge were included. Complete blood cell counts performed using a Sysmex XT-2000iV hematology analyzer were done on samples at room temperature (approximately 22°C) and following warming of the samples to 37°C in a water bath. The Wilcoxon signed rank test was used to compare hematologic variables, including platelet count, before and after sample warming to 37°C. Data are presented as median (25(th) , 75(th) percentile) increase 
520 |a RESULTS: Blood samples from 39 dogs, 19 cats, and 10 horses were included. Sample warming to 37°C resulted in a statistically significant increase in platelet counts in dogs (11 [-2, 30] ×10(9) /L), cats (36 [14, 84] ×10(9) /L), and horses (42 [31, 79] ×10(9) /L). Sample warming did not significantly affect other hematologic variables 
520 |a CONCLUSIONS: Prewarming EDTA blood samples to 37°C prior to hematologic analysis increased platelet counts overall in canine, feline, and equine blood, but did not abrogate platelet clumping and pseudothrombocytopenia fully in some cases. Furthermore, true pseudothrombocytopenia was not confirmed in these animals 
650 4 |a Journal Article 
650 4 |a Platelet clumping 
650 4 |a preanalytic error 
650 4 |a pseudothrombocytopenia 
650 7 |a Edetic Acid  |2 NLM 
650 7 |a 9G34HU7RV0  |2 NLM 
700 1 |a Archer, Joy  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Veterinary clinical pathology  |d 1975  |g 45(2016), 3 vom: 16. Sept., Seite 444-9  |w (DE-627)NLM098159984  |x 1939-165X  |7 nnns 
773 1 8 |g volume:45  |g year:2016  |g number:3  |g day:16  |g month:09  |g pages:444-9 