Validation of an automated enzyme immunoassay for the measurement of serum total thyroxine in cats

© 2016 American Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 45(2016), 1 vom: 01. März, Seite 148-53
1. Verfasser: Williams, Tim L (VerfasserIn)
Weitere Verfasser: Archer, Joy
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2016
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Journal Article Validation Study Analyzer feline human hyperthyroidism reference interval Thyroxine Q51BO43MG4
LEADER 01000caa a22002652c 4500
001 NLM257129685
003 DE-627
005 20250219161846.0
007 cr uuu---uuuuu
008 231224s2016 xx |||||o 00| ||eng c
024 7 |a 10.1111/vcp.12324  |2 doi 
028 5 2 |a pubmed25n0857.xml 
035 |a (DE-627)NLM257129685 
035 |a (NLM)26840919 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Williams, Tim L  |e verfasserin  |4 aut 
245 1 0 |a Validation of an automated enzyme immunoassay for the measurement of serum total thyroxine in cats 
264 1 |c 2016 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 28.02.2018 
500 |a Date Revised 10.12.2019 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a © 2016 American Society for Veterinary Clinical Pathology. 
520 |a BACKGROUND: Hyperthyroidism is common in older cats, which necessitates frequent screening of serum total thyroxine (TT4) concentrations. Fast, cheap, and reliable methods to measure TT4 in cats are needed 
520 |a OBJECTIVES: The purpose of the study was the validation of a human TT4 enzyme immunoassay (EIA) for use with feline serum, and derivation of a TT4 reference interval (RI) for cats aged 9 years and older 
520 |a METHODS: Assay precision, reproducibility, and linearity were evaluated. Interference by hemolysis was also assessed. Method comparison studies between the human EIA and a previously validated radioimmunoassay (RIA) and chemiluminescent-enzyme immunoassay (CEIA) were performed. Healthy cats (> 9 years) were recruited from 3 UK first opinion practices 
520 |a RESULTS: The human TT4 EIA demonstrated good precision and reproducibility, and adequate linearity. Hemolysis did not significantly alter measured TT4 concentrations until HGB > 8 g/L. Method comparison revealed proportional and constant errors between EIA and RIA/CEIA. The TT4 RI for cats (> 9 years) was calculated as 7.1-45.1 nmol/L (n = 49) 
520 |a CONCLUSIONS: The human TT4 EIA was successfully validated for use with feline serum and offers a rapid, cheap, and reliable method for determination of serum TT4 concentrations in cats 
650 4 |a Journal Article 
650 4 |a Validation Study 
650 4 |a Analyzer 
650 4 |a feline 
650 4 |a human 
650 4 |a hyperthyroidism 
650 4 |a reference interval 
650 7 |a Thyroxine  |2 NLM 
650 7 |a Q51BO43MG4  |2 NLM 
700 1 |a Archer, Joy  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Veterinary clinical pathology  |d 1975  |g 45(2016), 1 vom: 01. März, Seite 148-53  |w (DE-627)NLM098159984  |x 1939-165X  |7 nnas 
773 1 8 |g volume:45  |g year:2016  |g number:1  |g day:01  |g month:03  |g pages:148-53 
856 4 0 |u http://dx.doi.org/10.1111/vcp.12324  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 