A New Time Calibration Method for Switched-capacitor-array-based Waveform Samplers

We have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of a...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment. - 1987. - 767(2014) vom: 11. Dez., Seite 67-74
1. Verfasser: Kim, H (VerfasserIn)
Weitere Verfasser: Chen, C-T, Eclov, N, Ronzhin, A, Murat, P, Ramberg, E, Los, S, Moses, W, Choong, W-S, Kao, C-M
Format: Aufsatz
Sprache:English
Veröffentlicht: 2014
Zugriff auf das übergeordnete Werk:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment
Schlagworte:Journal Article Switched-Capacitors Time Calibration Waveform Sampling
LEADER 01000naa a22002652 4500
001 NLM244581908
003 DE-627
005 20231224135019.0
007 tu
008 231224s2014 xx ||||| 00| ||eng c
028 5 2 |a pubmed24n0815.xml 
035 |a (DE-627)NLM244581908 
035 |a (NLM)25506113 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Kim, H  |e verfasserin  |4 aut 
245 1 2 |a A New Time Calibration Method for Switched-capacitor-array-based Waveform Samplers 
264 1 |c 2014 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Revised 20.11.2019 
500 |a published: Print 
500 |a Citation Status Publisher 
520 |a We have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibration is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. The new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration 
650 4 |a Journal Article 
650 4 |a Switched-Capacitors 
650 4 |a Time Calibration 
650 4 |a Waveform Sampling 
700 1 |a Chen, C-T  |e verfasserin  |4 aut 
700 1 |a Eclov, N  |e verfasserin  |4 aut 
700 1 |a Ronzhin, A  |e verfasserin  |4 aut 
700 1 |a Murat, P  |e verfasserin  |4 aut 
700 1 |a Ramberg, E  |e verfasserin  |4 aut 
700 1 |a Los, S  |e verfasserin  |4 aut 
700 1 |a Moses, W  |e verfasserin  |4 aut 
700 1 |a Choong, W-S  |e verfasserin  |4 aut 
700 1 |a Kao, C-M  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment  |d 1987  |g 767(2014) vom: 11. Dez., Seite 67-74  |w (DE-627)NLM098171771  |x 0168-9002  |7 nnns 
773 1 8 |g volume:767  |g year:2014  |g day:11  |g month:12  |g pages:67-74 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 767  |j 2014  |b 11  |c 12  |h 67-74