Acoustic characterization of contrast-to-tissue ratio and axial resolution for dual-frequency contrast-specific acoustic angiography imaging
Recently, dual-frequency transducers have enabled high-spatial-resolution and high-contrast imaging of vasculature with minimal tissue artifacts by transmitting at a low frequency and receiving broadband superharmonic echoes scattered by microbubble contrast agents. In this work, we examine the imag...
| Publié dans: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 61(2014), 10 vom: 22. Okt., Seite 1668-87 |
|---|---|
| Auteur principal: | |
| Autres auteurs: | , , , |
| Format: | Article en ligne |
| Langue: | English |
| Publié: |
2014
|
| Accès à la collection: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control |
| Sujets: | Journal Article Research Support, N.I.H., Extramural Contrast Media |
| Résumé: | Recently, dual-frequency transducers have enabled high-spatial-resolution and high-contrast imaging of vasculature with minimal tissue artifacts by transmitting at a low frequency and receiving broadband superharmonic echoes scattered by microbubble contrast agents. In this work, we examine the imaging parameters for optimizing contrast-to-tissue ratio (CTR) for dual-frequency imaging and the relationship with spatial resolution. Confocal piston transducers are used in a water bath setup to measure the SNR, CTR, and axial resolution for ultrasound imaging of nonlinear scattering of microbubble contrast agents when transmitting at a lower frequency (1.5 to 8 MHz) and receiving at a higher frequency (7.5 to 25 MHz). Parameters varied include the frequency and peak negative pressure of transmitted waves, center frequency of the receiving transducer, microbubble concentration, and microbubble size. CTR is maximized at the lowest transmission frequencies but would be acceptable for imaging in the 1.5 to 3.5 MHz range. At these frequencies, CTR is optimized when a receiving transducer with a center frequency of 10 MHz is used, with the maximum CTR of 25.5 dB occurring when transmitting at 1.5 MHz with a peak negative pressure of 1600 kPa and receiving with a center frequency of 10 MHz. Axial resolution is influenced more heavily by the receiving center frequency, with a weak decrease in measured pulse lengths associated with increasing transmit frequency. A microbubble population containing predominately 4-μm-diameter bubbles yielded the greatest CTR, followed by 1- and then 2-μm bubbles. Varying concentration showed little effect over the tested parameters. CTR dependence on transmit frequency and peak pressure were confirmed through in vivo imaging in two rodents. These findings may lead to improved imaging of vascular remodeling in superficial or luminal cancers such as those of the breast, prostate, and colon |
|---|---|
| Description: | Date Completed 26.05.2015 Date Revised 21.10.2021 published: Print Citation Status MEDLINE |
| ISSN: | 1525-8955 |
| DOI: | 10.1109/TUFFC.2014.006466 |