Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines

OBJECTIVE: To assure the accuracy and reliability of blood pressure measurement by non-invasive blood pressure monitor using Datascope Accutorr Plus™ (Paramus, NJ, USA) against mercury manometer, among adult male participants

Bibliographische Detailangaben
Veröffentlicht in:Journal of the Saudi Heart Association. - 1999. - 22(2010), 1 vom: 01. Jan., Seite 1-5
1. Verfasser: Khawaja, Rajab Ali (VerfasserIn)
Weitere Verfasser: Qureshi, Riaz, Mansure, Abdul Hay, Yahya, Mohammad Emran
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2010
Zugriff auf das übergeordnete Werk:Journal of the Saudi Heart Association
Schlagworte:Journal Article AAMI BHS BP devices Datascope Acutorr plus Validation
LEADER 01000naa a22002652 4500
001 NLM239248325
003 DE-627
005 20231224115531.0
007 cr uuu---uuuuu
008 231224s2010 xx |||||o 00| ||eng c
024 7 |a 10.1016/j.jsha.2010.03.001  |2 doi 
028 5 2 |a pubmed24n0797.xml 
035 |a (DE-627)NLM239248325 
035 |a (NLM)24936124 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Khawaja, Rajab Ali  |e verfasserin  |4 aut 
245 1 0 |a Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines 
264 1 |c 2010 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 17.06.2014 
500 |a Date Revised 04.11.2023 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a OBJECTIVE: To assure the accuracy and reliability of blood pressure measurement by non-invasive blood pressure monitor using Datascope Accutorr Plus™ (Paramus, NJ, USA) against mercury manometer, among adult male participants 
520 |a METHOD: Eighty participants from a family physician's office at a teaching hospital were recruited. One hundred and sixty measurements of blood pressure were performed according to BHS technique protocol 
520 |a STATISTICAL ANALYSIS: Descriptive analysis was done according to the AAMI and BHS protocol guidelines. The limits of agreement between the device and the standard were plotted using the method of Bland and Altman plot 
520 |a RESULTS: The mean difference ± SD between the Datascope Acutorr Plus™ and observer was 2.7 ± 5.2 mm Hg and 1.5 ± 3.26 mm Hg for systolic and diastolic blood pressure, respectively. Datascope Acutorr Plus™ obtained A/A grading for both systolic and diastolic blood pressure 
520 |a CONCLUSION: Datascope Accutorr Plus™ (Paramus, NJ, USA) satisfies BHS and AAMI validation protocols for both systolic and diastolic BP and may be recommended for everyday use for BP monitoring at home and in clinical use for adult population 
650 4 |a Journal Article 
650 4 |a AAMI 
650 4 |a BHS 
650 4 |a BP devices 
650 4 |a Datascope Acutorr plus 
650 4 |a Validation 
700 1 |a Qureshi, Riaz  |e verfasserin  |4 aut 
700 1 |a Mansure, Abdul Hay  |e verfasserin  |4 aut 
700 1 |a Yahya, Mohammad Emran  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of the Saudi Heart Association  |d 1999  |g 22(2010), 1 vom: 01. Jan., Seite 1-5  |w (DE-627)NLM098225227  |x 1016-7315  |7 nnns 
773 1 8 |g volume:22  |g year:2010  |g number:1  |g day:01  |g month:01  |g pages:1-5 
856 4 0 |u http://dx.doi.org/10.1016/j.jsha.2010.03.001  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 22  |j 2010  |e 1  |b 01  |c 01  |h 1-5