ASVCP guidelines : quality assurance for point-of-care testing in veterinary medicine

© 2013 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 42(2013), 4 vom: 23. Dez., Seite 405-23
1. Verfasser: Flatland, Bente (VerfasserIn)
Weitere Verfasser: Freeman, Kathleen P, Vap, Linda M, Harr, Kendal E, ASVCP
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2013
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Journal Article Practice Guideline Allowable total error bench-top handheld in-clinic near-patient quality control
LEADER 01000naa a22002652 4500
001 NLM233496165
003 DE-627
005 20231224095233.0
007 cr uuu---uuuuu
008 231224s2013 xx |||||o 00| ||eng c
024 7 |a 10.1111/vcp.12099  |2 doi 
028 5 2 |a pubmed24n0778.xml 
035 |a (DE-627)NLM233496165 
035 |a (NLM)24320778 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Flatland, Bente  |e verfasserin  |4 aut 
245 1 0 |a ASVCP guidelines  |b quality assurance for point-of-care testing in veterinary medicine 
264 1 |c 2013 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 12.05.2015 
500 |a Date Revised 20.10.2016 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a © 2013 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology. 
520 |a Point-of-care testing (POCT) refers to any laboratory testing performed outside the conventional reference laboratory and implies close proximity to patients. Instrumental POCT systems consist of small, handheld or benchtop analyzers. These have potential utility in many veterinary settings, including private clinics, academic veterinary medical centers, the community (eg, remote area veterinary medical teams), and for research applications in academia, government, and industry. Concern about the quality of veterinary in-clinic testing has been expressed in published veterinary literature; however, little guidance focusing on POCT is available. Recognizing this void, the ASVCP formed a subcommittee in 2009 charged with developing quality assurance (QA) guidelines for veterinary POCT. Guidelines were developed through literature review and a consensus process. Major recommendations include (1) taking a formalized approach to POCT within the facility, (2) use of written policies, standard operating procedures, forms, and logs, (3) operator training, including periodic assessment of skills, (4) assessment of instrument analytical performance and use of both statistical quality control and external quality assessment programs, (5) use of properly established or validated reference intervals, (6) and ensuring accurate patient results reporting. Where possible, given instrument analytical performance, use of a validated 13s control rule for interpretation of control data is recommended. These guidelines are aimed at veterinarians and veterinary technicians seeking to improve management of POCT in their clinical or research setting, and address QA of small chemistry and hematology instruments. These guidelines are not intended to be all-inclusive; rather, they provide a minimum standard for maintenance of POCT instruments in the veterinary setting 
650 4 |a Journal Article 
650 4 |a Practice Guideline 
650 4 |a Allowable total error 
650 4 |a bench-top 
650 4 |a handheld 
650 4 |a in-clinic 
650 4 |a near-patient 
650 4 |a quality control 
700 1 |a Freeman, Kathleen P  |e verfasserin  |4 aut 
700 1 |a Vap, Linda M  |e verfasserin  |4 aut 
700 1 |a Harr, Kendal E  |e verfasserin  |4 aut 
700 0 |a ASVCP  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Veterinary clinical pathology  |d 1975  |g 42(2013), 4 vom: 23. Dez., Seite 405-23  |w (DE-627)NLM098159984  |x 1939-165X  |7 nnns 
773 1 8 |g volume:42  |g year:2013  |g number:4  |g day:23  |g month:12  |g pages:405-23 
856 4 0 |u http://dx.doi.org/10.1111/vcp.12099  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 42  |j 2013  |e 4  |b 23  |c 12  |h 405-23