Comparison of the buccal mucosal bleeding time in dogs using 3 different-sized lancet devices

© 2013 American Society for Veterinary Clinical Pathology.

Bibliographische Detailangaben
Veröffentlicht in:Veterinary clinical pathology. - 1975. - 42(2013), 4 vom: 11. Dez., Seite 451-7
1. Verfasser: Aumann, Marcel (VerfasserIn)
Weitere Verfasser: Rossi, Valentina, Le Boedec, Kevin, Diquélou, Armelle
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2013
Zugriff auf das übergeordnete Werk:Veterinary clinical pathology
Schlagworte:Comparative Study Journal Article Dogs hemorrhage mucosa platelets primary hemostasis
Beschreibung
Zusammenfassung:© 2013 American Society for Veterinary Clinical Pathology.
BACKGROUND: The buccal mucosal bleeding time (BMBT) evaluates primary hemostasis in vivo. Three different-sized lancet devices designed for people, Adult (A), Junior (J), and Newborn (N), can be used to perform the BMBT in dogs
OBJECTIVES: The aim was to compare BMBT using 3 different-sized human lancet devices in dogs with varying platelet counts and hematocrits
METHODS: The BMBT was measured in 46 client-owned dogs (2 healthy, 44 suffering from various disorders) with varying platelet (Plt) counts and hematocrits, using the 3 devices successively in each dog, in a randomly determined order, over a 10- to 30-minute period. Statistical analysis (ANOVA, Mann-Whitney U-test) was performed using commercial software
RESULTS: BMBTs were significantly different between devices (P < .00001), and shorter with devices N and J compared with device A (P < .01). The BMBT was prolonged (> 210 s) in 10 dogs with device A and in 7 dogs each with devices J and N, respectively. Sixteen dogs had a Plt count < 200 × 10(9) /L (Reference interval 200-500 × 10(9) /L). Nine of these dogs had prolonged BMBT with device A, and 6 dogs with device J and device N, respectively. BMBT was longer in thrombocytopenic dogs with devices A and J (P < .016). Anemia without thrombocytopenia did not affect BMBT with any device
CONCLUSIONS: The BMBT is influenced by the size of the used device, with A resulting in the longest BMBT. Therefore, the type of device used to obtain the BMBT has to be specified for standardized results