Success and failure of dead-time models as applied to hybrid pixel detectors in high-flux applications

The performance of a single-photon-counting hybrid pixel detector has been investigated at the Australian Synchrotron. Results are compared with the body of accepted analytical models previously validated with other detectors. Detector functionals are valuable for empirical calibration. It is shown...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 20(2013), Pt 2 vom: 15. März, Seite 347-54
1. Verfasser: Sobott, B A (VerfasserIn)
Weitere Verfasser: Broennimann, Ch, Schmitt, B, Trueb, P, Schneebeli, M, Lee, V, Peake, D J, Elbracht-Leong, S, Schubert, A, Kirby, N, Boland, M J, Chantler, C T, Barnea, Z, Rassool, R P
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2013
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article Research Support, Non-U.S. Gov't dead-time hybrid pixel detector single-photon counting synchrotron fill pattern
LEADER 01000caa a22002652 4500
001 NLM225054450
003 DE-627
005 20241006231809.0
007 cr uuu---uuuuu
008 231224s2013 xx |||||o 00| ||eng c
024 7 |a 10.1107/S0909049513000411  |2 doi 
028 5 2 |a pubmed24n1559.xml 
035 |a (DE-627)NLM225054450 
035 |a (NLM)23412493 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Sobott, B A  |e verfasserin  |4 aut 
245 1 0 |a Success and failure of dead-time models as applied to hybrid pixel detectors in high-flux applications 
264 1 |c 2013 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 06.08.2013 
500 |a Date Revised 06.10.2024 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a The performance of a single-photon-counting hybrid pixel detector has been investigated at the Australian Synchrotron. Results are compared with the body of accepted analytical models previously validated with other detectors. Detector functionals are valuable for empirical calibration. It is shown that the matching of the detector dead-time with the temporal synchrotron source structure leads to substantial improvements in count rate and linearity of response. Standard implementations are linear up to ∼0.36 MHz pixel(-1); the optimized linearity in this configuration has an extended range up to ∼0.71 MHz pixel(-1); these are further correctable with a transfer function to ∼1.77 MHz pixel(-1). This new approach has wide application both in high-accuracy fundamental experiments and in standard crystallographic X-ray fluorescence and other X-ray measurements. The explicit use of data variance (rather than N(1/2) noise) and direct measures of goodness-of-fit (χ(r)(2)) are introduced, raising issues not encountered in previous literature for any detector, and suggesting that these inadequacies of models may apply to most detector types. Specifically, parametrization of models with non-physical values can lead to remarkable agreement for a range of count-rate, pulse-frequency and temporal structure. However, especially when the dead-time is near resonant with the temporal structure, limitations of these classical models become apparent. Further, a lack of agreement at extreme count rates was evident 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
650 4 |a dead-time 
650 4 |a hybrid pixel detector 
650 4 |a single-photon counting 
650 4 |a synchrotron fill pattern 
700 1 |a Broennimann, Ch  |e verfasserin  |4 aut 
700 1 |a Schmitt, B  |e verfasserin  |4 aut 
700 1 |a Trueb, P  |e verfasserin  |4 aut 
700 1 |a Schneebeli, M  |e verfasserin  |4 aut 
700 1 |a Lee, V  |e verfasserin  |4 aut 
700 1 |a Peake, D J  |e verfasserin  |4 aut 
700 1 |a Elbracht-Leong, S  |e verfasserin  |4 aut 
700 1 |a Schubert, A  |e verfasserin  |4 aut 
700 1 |a Kirby, N  |e verfasserin  |4 aut 
700 1 |a Boland, M J  |e verfasserin  |4 aut 
700 1 |a Chantler, C T  |e verfasserin  |4 aut 
700 1 |a Barnea, Z  |e verfasserin  |4 aut 
700 1 |a Rassool, R P  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 20(2013), Pt 2 vom: 15. März, Seite 347-54  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:20  |g year:2013  |g number:Pt 2  |g day:15  |g month:03  |g pages:347-54 
856 4 0 |u http://dx.doi.org/10.1107/S0909049513000411  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 20  |j 2013  |e Pt 2  |b 15  |c 03  |h 347-54