Automatic loop centring with a high-precision goniometer head at the SLS macromolecular crystallography beamlines

Automatic loop centring has been developed as part of the automation process in crystallographic data collection at the Swiss Light Source. The procedure described here consists of an optional set-up part, in which the background images are taken, and the actual centring part. The algorithm uses bou...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 18(2011), Pt 4 vom: 22. Juli, Seite 595-600
1. Verfasser: Pauluhn, Anuschka (VerfasserIn)
Weitere Verfasser: Pradervand, Claude, Rossetti, Daniel, Salathe, Marco, Schulze-Briese, Clemens
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2011
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
LEADER 01000naa a22002652 4500
001 NLM20928479X
003 DE-627
005 20231224005512.0
007 cr uuu---uuuuu
008 231224s2011 xx |||||o 00| ||eng c
024 7 |a 10.1107/S0909049511011848  |2 doi 
028 5 2 |a pubmed24n0698.xml 
035 |a (DE-627)NLM20928479X 
035 |a (NLM)21685676 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Pauluhn, Anuschka  |e verfasserin  |4 aut 
245 1 0 |a Automatic loop centring with a high-precision goniometer head at the SLS macromolecular crystallography beamlines 
264 1 |c 2011 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 13.10.2011 
500 |a Date Revised 05.11.2023 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a Automatic loop centring has been developed as part of the automation process in crystallographic data collection at the Swiss Light Source. The procedure described here consists of an optional set-up part, in which the background images are taken, and the actual centring part. The algorithm uses boundary and centre-of-mass detection at two different microscope image magnifications. Micromounts can be handled as well. Centring of the loops can be achieved in 15-26s, depending on their initial position, and as fast as manual centring. The alignment of the sample is carried out by means of a new flexural-hinge-based compact goniometer head. The device features an electromagnet for robotic wet mounting of samples. The circle of confusion was measured to be smaller than 1 µm (r.m.s.); its bidirectional backlash is below 2 µm 
650 4 |a Journal Article 
700 1 |a Pradervand, Claude  |e verfasserin  |4 aut 
700 1 |a Rossetti, Daniel  |e verfasserin  |4 aut 
700 1 |a Salathe, Marco  |e verfasserin  |4 aut 
700 1 |a Schulze-Briese, Clemens  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 18(2011), Pt 4 vom: 22. Juli, Seite 595-600  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:18  |g year:2011  |g number:Pt 4  |g day:22  |g month:07  |g pages:595-600 
856 4 0 |u http://dx.doi.org/10.1107/S0909049511011848  |3 Volltext 