Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography

X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 18(2011), Pt 2 vom: 21. März, Seite 217-23
1. Verfasser: Uesugi, K (VerfasserIn)
Weitere Verfasser: Hoshino, M, Yagi, N
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2011
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM20600463X
003 DE-627
005 20250212114413.0
007 cr uuu---uuuuu
008 231223s2011 xx |||||o 00| ||eng c
024 7 |a 10.1107/S0909049510044523  |2 doi 
028 5 2 |a pubmed25n0687.xml 
035 |a (DE-627)NLM20600463X 
035 |a (NLM)21335908 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Uesugi, K  |e verfasserin  |4 aut 
245 1 0 |a Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography 
264 1 |c 2011 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 03.06.2011 
500 |a Date Revised 20.10.2021 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It was found that the transmittance of light to the CCD is about four times higher in the fiber-coupled detector. The uniformity of response in the lens-coupled detector has a global shading of up to 40%, while pixel-to-pixel variation owing to a chicken-wire pattern was dominant in the fiber-coupled detector. Apart from the higher transmittance, the fiber-coupled detector has a few characteristics that require attention when it is used for computed tomography, which are browning of the fiber, discontinuity in the image, image distortion, and dark spots in the chicken-wire pattern. Thus, it is most suitable for high-speed tomography of samples that tend to deform, for example biological and soft materials 
650 4 |a Journal Article 
700 1 |a Hoshino, M  |e verfasserin  |4 aut 
700 1 |a Yagi, N  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 18(2011), Pt 2 vom: 21. März, Seite 217-23  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:18  |g year:2011  |g number:Pt 2  |g day:21  |g month:03  |g pages:217-23 
856 4 0 |u http://dx.doi.org/10.1107/S0909049510044523  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 18  |j 2011  |e Pt 2  |b 21  |c 03  |h 217-23