A high-frequency, high frame rate duplex ultrasound linear array imaging system for small animal imaging

High-frequency (HF) ultrasound imaging has been shown to be useful for non-invasively imaging anatomical structures of the eye and small animals in biological and pharmaceutical research, achieving superior spatial resolution. Cardiovascular research utilizing mice requires not only realtime B-scan...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 57(2010), 7 vom: 01. Juli, Seite 1548-57
1. Verfasser: Zhang, Lequan (VerfasserIn)
Weitere Verfasser: Xu, Xiaochen, Hu, Changhong, Sun, Lei, Yen, Jesse T, Cannata, Jonathan M, Shung, K Kirk
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2010
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, N.I.H., Extramural
LEADER 01000naa a22002652 4500
001 NLM199659672
003 DE-627
005 20231223215101.0
007 cr uuu---uuuuu
008 231223s2010 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2010.1585  |2 doi 
028 5 2 |a pubmed24n0666.xml 
035 |a (DE-627)NLM199659672 
035 |a (NLM)20639149 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Zhang, Lequan  |e verfasserin  |4 aut 
245 1 2 |a A high-frequency, high frame rate duplex ultrasound linear array imaging system for small animal imaging 
264 1 |c 2010 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 19.01.2011 
500 |a Date Revised 20.10.2021 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a High-frequency (HF) ultrasound imaging has been shown to be useful for non-invasively imaging anatomical structures of the eye and small animals in biological and pharmaceutical research, achieving superior spatial resolution. Cardiovascular research utilizing mice requires not only realtime B-scan imaging, but also ultrasound Doppler to evaluate both anatomy and blood flow of the mouse heart. This paper reports the development of an HF ultrasound duplex imaging system capable of both B-mode imaging and Doppler flow measurements, using a 64-element linear array. The system included an HF pulsed-wave Doppler module, a 32-channel HF B-mode imaging module, a PC with a 200 MS/s 14-bit A/D card, and real-time LabView software. A 50 dB SNR and a depth of penetration of larger than 12 mm were achieved using a 35-MHz linear array with 50 μm pitch. The two-way beam widths were determined to be 165 to 260 μm and the clutter-energy-to-total-energy ratio (CTR) were 9.1 to 12 dB when the array was electronically focused at different focal points at depths from 4.8 to 9.6 mm. The system is capable of acquiring real-time B-mode images at a rate greater than 400 frames per second (fps) for a 4.8 x 13 mm field of view, using a 30- MHz 64-element linear array with 100 μm pitch. Sample in vivo cardiac high frame rate images and duplex images of mouse hearts are shown to assess its current imaging capability and performance for small animals 
650 4 |a Journal Article 
650 4 |a Research Support, N.I.H., Extramural 
700 1 |a Xu, Xiaochen  |e verfasserin  |4 aut 
700 1 |a Hu, Changhong  |e verfasserin  |4 aut 
700 1 |a Sun, Lei  |e verfasserin  |4 aut 
700 1 |a Yen, Jesse T  |e verfasserin  |4 aut 
700 1 |a Cannata, Jonathan M  |e verfasserin  |4 aut 
700 1 |a Shung, K Kirk  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 57(2010), 7 vom: 01. Juli, Seite 1548-57  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:57  |g year:2010  |g number:7  |g day:01  |g month:07  |g pages:1548-57 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2010.1585  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 57  |j 2010  |e 7  |b 01  |c 07  |h 1548-57