Influence of pulse indicator continuous cardiac output in monitoring on the volume parameters of different tidal volume under pressure control ventilation mode and volume-controlled ventilation mode in sheep

OBJECTIVE: To compare the influence of pulse indicator continuous cardiac output (PiCCO) in monitoring tidal volume (V(T)) under pressure control ventilation mode and volume-controlled ventilation mode in sheep

Bibliographische Detailangaben
Veröffentlicht in:Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue. - 1998. - 22(2010), 3 vom: 06. März, Seite 146-9
1. Verfasser: Li, Jun (VerfasserIn)
Weitere Verfasser: Gao, Xin-jing, Gao, Yan-ying, Wei, Jun-tao, Qin, Ying-zhi
Format: Aufsatz
Sprache:Chinese
Veröffentlicht: 2010
Zugriff auf das übergeordnete Werk:Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue
Schlagworte:Comparative Study English Abstract Journal Article Research Support, Non-U.S. Gov't
LEADER 01000caa a22002652 4500
001 NLM197120628
003 DE-627
005 20250211105207.0
007 tu
008 231223s2010 xx ||||| 00| ||chi c
028 5 2 |a pubmed25n0657.xml 
035 |a (DE-627)NLM197120628 
035 |a (NLM)20367903 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a chi 
100 1 |a Li, Jun  |e verfasserin  |4 aut 
245 1 0 |a Influence of pulse indicator continuous cardiac output in monitoring on the volume parameters of different tidal volume under pressure control ventilation mode and volume-controlled ventilation mode in sheep 
264 1 |c 2010 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 07.12.2010 
500 |a Date Revised 06.04.2010 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a OBJECTIVE: To compare the influence of pulse indicator continuous cardiac output (PiCCO) in monitoring tidal volume (V(T)) under pressure control ventilation mode and volume-controlled ventilation mode in sheep 
520 |a METHODS: After anesthesia and tracheotomy, 5 sheep, which were apneic and receiving mechanical ventilation. Twenty minutes later, central venous pressure (CVP) and cardiac function were monitored with different selected V(T) levels of 6, 10, 15, 20 ml/kg under bi-level positive airway pressure (BiPAP) mode by changing the pressure of inspiration, or under the synchronized intermittent mandatory ventilation (SIMV) mode with the same ventilation conditions 
520 |a RESULTS: In both modes, the increase in V(T) led to an decrease of cardiac index (CI) and intrathoracic blood volume index (ITBVI), reaching a statistically significant difference at 15 ml/kg [SIMV mode: CI (3.94 + or - 1.03) L x min(-1) x m(-2), ITBVI (707 + or - 105) ml/m(2); BiPAP mode: CI (4.11 + or - 1.11) L x mi(-1) x m(-2), ITBVI (715 + or - 122) ml/m(2)] and 20 ml/kg [SIMV mode: CI (3.87 + or - 1.04) L x min(-1) x m(-2), ITBVI (705 + or - 116) ml/m(2); BiPAP mode: CI (3.64 + or - 0.96) L x min(-1) x m(-2), ITBVI (694 + or - 114) ml/m(2)] compared with 6 ml/kg [SIMV mode: CI (4.96 + or - 1.58) L x min(-1) x m(-2),ITBVI(811 + or - 169) ml/m(2); BiPAP mode:CI(5.67 + or - 1.96) L x min(-1) x m(-2), ITBVI (823 + or - 182) ml/m(2), all P<0.05]; an increase in systemic vascular resistance index (SVRI) and mean airway pressure (Pmean) at 15 ml/kg [SIMV mode: SVRI (237.6 + or - 56.2) kPaxs(-1) x L(-1), Pmean (14.0 + or - 3.2) cm H(2)O (1 cm H(2)O=0.098 kPa); BiPAP mode: SVRI (230.8 + or - 32.9) kPaxs(-1) x L(-1), Pmean (13.0 + or - 2.2) cm H(2)O] and 20 ml/kg [SIMV mode: SVRI (253.1 + or - 76.7) kPaxs(-1) x L(-1), Pmean (18.2 + or - 4.8) cm H(2)O ; BiPAP mode: SVRI (246.7 + or - 48.8) kPaxs(-1) x L(-1), Pmean (16.8 + or - 3.3) cm H(2)O] compared with 6 ml/kg [SIMV mode: SVRI (184.8 + or - 47.5) kPaxs(-1) x L(-1); Pmean (8.8 + or - 1.6) cm H(2)O; BiPAP mode: SVRI (184.5 + or - 51.5) kPaxs(-1) x L(-1), Pmean (8.6 + or - 0.5) cm H(2)O, all P<0.05]; but there was no significant effects on CVP, heart rate (HR), mean blood pressure (MBP). There was no significant difference of CI, ITBVI, SVRI and Pmean between the two ventilation modes with various V(T) levels 
520 |a CONCLUSION: When the cardiac function was normal, the increase in V(T) led to a decrease of CI and ITBVI, but it had no significant effects on CVP. There was no significant difference of CI and ITBVI in the two ventilation modes, both were decreased. So a relatively constant V(T) should be maintained in determining ITBVI 
650 4 |a Comparative Study 
650 4 |a English Abstract 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
700 1 |a Gao, Xin-jing  |e verfasserin  |4 aut 
700 1 |a Gao, Yan-ying  |e verfasserin  |4 aut 
700 1 |a Wei, Jun-tao  |e verfasserin  |4 aut 
700 1 |a Qin, Ying-zhi  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue  |d 1998  |g 22(2010), 3 vom: 06. März, Seite 146-9  |w (DE-627)NLM098227793  |x 1003-0603  |7 nnns 
773 1 8 |g volume:22  |g year:2010  |g number:3  |g day:06  |g month:03  |g pages:146-9 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 22  |j 2010  |e 3  |b 06  |c 03  |h 146-9